Note
•
If the probe check result does not show a value of 1.9 - 2.1 mmho/cc/cm
3
/ml at 226 Hz, we recommend that
you make an admittance calibration. See the Zodiac Reference Manual.
Probe home
You can mount the probe home on the wall, using the optional probe home wall-mount kit.
Powering
•
See
.
4.1
Powering the device
Zodiac is powered through an external power supply connected directly to the mains outlet.
Caution
•
Use only the power supply specified in
Connecting the external power supply to Zodiac
1. Connect the plug end of the external power supply cable to the external power supply socket on the
back of the device.
Connecting the external power supply to the mains supply
1. Connect the mains plug of the external power supply directly to an AC mains outlet with a three-wire
protective ground.
2. If applicable, switch on the mains supply.
The first time you switch on the device
Note
•
The first time you switch on the device, leave it turned on for at least an hour to let the internal clock battery
charge.
The first time you switch on the device, or if the device has been switched off for more than two weeks, the internal clock
runs out of power. When you start up the device, you will be prompted to set the time manually.
•
Set the date and time as required on the device.
•
Alternatively, you can connect to OTOsuite where it will be done automatically.
7
User Guide
Otometrics - MADSEN Zodiac