LBN
[3]
7.5
9.5
8.5
10.5
11.5
HBN
[3]
4.0
5.0
7.5
12.5
14.5
Notes:
1. From ANSI/ASA S3.6-2010, Table 5.
2. From ANSI/ASA S3.6-2010, Table 7.
3. Based on Otometrics internal study
Air pressure system
Range:
200 to -400 daPa/s
Pressure sweep rate:
200, 400, 600 daPa/s ± 20% in 20% to 80% of the total pressure range
Pressure accuracy:
± 10% or ± 10 daPa, whichever is greatest
Pump measure direction:
Positive to negative or negative to positive
Safety:
Separate 530 daPa and -730 daPa ±70 daPa
Graph units
Unit of admittance graph Y-axis:
ml, cc, mmho, µl
Unit of graph X-axis:
daPa, sec
Device display
Display:
7 inch, 15:9 WVGA
Resolution:
800 x 480 pixel
USB port connector
Type:
USB device port
Power supply
External power supply
XP Power, type AFM60US24
Output:
24 V, 2.5 A
Input:
100-240 V AC, 50-60 Hz, 1.5 A
Power consumption
Power consumption:
< 70 VA
Operating environment
Temperature:
+15°C to +35°C (59°F to +95°F)
Air humidity:
10 to 90%, non-condensing
29
User Guide
Otometrics - MADSEN Zodiac