10.6 PRINTER
Your MIDIMATE 622 is equipped with a printout capability. This facility
enables Air, Bone and Free-Field data for both ears to be memorized and
printed out as audiograms to:
MIDIMATE 622
Printout Capability
a standard external printer (serial interface)
ZODIAC 901's built-in printer (serial interface)
ZODIAC 901's external printer (parallel interface)
Numerous setup options let you program the instrument for the following
parameters:
!
Hardware connection: baud rate and handshake
!
Header: Madsen's logo and patient I.D. data
!
Select printer: ZODIAC 901 or standard external printer
!
Format: 110 mm (ZODIAC 901), A4: L,R or A4: R,L
!
Threshold logging: each time a threshold is stored, data is printed out in
tabular form
!
Averaging values (with weighting) or masking shown on printout
A Madsen configuration program enables the definition of: form feeds, left
margin, line feeds, weighting factors, and printer driver.
Madsen Config.
Program
10.6.1 PRINTER OPTION MENU
Select Printer by pressing Softkey 5 and the following menu will be
displayed:
SOFTKEY 5
Printer Options: 105
Select
Format
Header Ths.Log Avg.
All printer setup parameters can be selected from this sub-menu.
After selecting one of the 5 choices and setting up your printer, return to the
Printer Option Menu by pressing
MENU
.
MENU
It should be noted that the current printer setup status is not evident in any
of the 5 sub-menus. However, the current status can be deduced from the
number in the top line of the Printer Option Menu. Please refer to Appendix
D for further details.
Current Setup
Status Table In
Appendix D
Pressing
MENU
while the display shows the Power-On Options Menu
returns the audiometer to normal operation.
Press
MENU
to Exit
MIDIMATE 622 OPERATION MANUAL
10. Power-On Options
60