During the SISI test, this key toggles between 1 & 5 dB intensity modulation.
SISI Test
Softkey 4
In Test Mode, Softkey 4 is used to switch masking on and off. The current
mode is indicated by a cursor.
Softkey 4
Softkey 5
In Setup Mode, Softkey 5 can be used to select Patient (I.D. # entry) from the
Select Function Menu.
Please refer to chapter 13 for further details concerning Patient I.D. entry.
Patient I.D. Entry
5.1.5 OPTION SOFTKEY
Pressing Softkey 3 while in Test Mode presents the available options for that
test currently selected on the display.
Softkey 3
NOTE
: If HLL-Off or PTS-Off is displayed, then these optional functions are
both
ON
and can be disabled by pressing the appropriate softkey. And vice
versa. Remember that the Options Menu is displaying whatever options
are available and
NOT
the status of these functions!
ATTENTION!
The following options exist from the Options Menu (no Power-On Options
enabled):
Options Menu:
1.5s On
1 dB
Datrf.
HLL-Off PTS-On
When the Printer Function has been enabled, Datrf. changes to Store:
Options Menu:
1.5s On
1 dB
Store*
HLL-Off PTS-On
When User-programming has been enabled, Store Changes to Progr.
Options Menu:
1.5s On
1 dB
Progr.** HLL-Off PTS-On
Store
*This option is only present when printout has been enabled. (See 10.
Power-On Options for details about enabling printout).
Printer Function
The purpose of this option is to enable the user to store the current threshold
for subsequent printout. However, if the softkey is held down for longer than
0.8 seconds, the Printer Menu appears enabling previously stored data to be
cleared, viewed or printed out.
Printer Menu
Holding down Softkey 3 accesses the Printer Menu:
Printer Menu
Printer Menu:
Clr-L
Clr-R
View
Plot
Clr-All
MIDIMATE 622 OPERATION MANUAL
5. Operating in Test Mode
26