To fix the hearing instrument on the adapter, snap the adapter off the coupler cavity, and attach the hearing instrument to
the adapter outside the test chamber.
3.3
Traditional BTE hearing instruments
This procedure applies to any type of standard BTE hearing instruments with traditional ear molds.
Positioning the hearing instrument for directional testing is described in
General guidelines for positioning the hearing
.
3.3.1
With BTE adapter tube
1. Attach the BTE adapter tube to the tube stub on the HA-2 BTE adapter.
The tube must completely cover the tube stub on the adapter.
If a metal wire protrudes from the BTE adapter tube, turn the tube
upside down or use a new one.
2. Attach the hearing instrument to the free end of the BTE adapter tube
so that the metal wires in the tube align with the sides of the hearing
instrument.
3. Click the adapter with hearing instrument into place on the coupler and
place the coupler in the low position in AURICAL HIT.
.
20
Otometrics - AURICAL HIT
3 Testing hearing instruments