ABP Report Management System Operations Manual
6-7
System Setup
3
Define the statistical period boundaries.
•
Enter the start time and end time for each period, as well as the expected average systolic
and diastolic blood pressure. Select
Sleep
if the period is during sleep.
•
The pie chart in the dialog box reflects the period settings. If the period is a sleep period,
an “S” will appear in the corresponding section of the pie chart.
4
Click
OK
to close the dialog box when you finish customizing the statistics settings, or select
the
Monitor
tab to continue.
Table 1: Edit Limits
Systolic Max
Highest acceptable value for a systolic reading. The value entered cannot exceed
250 and must be greater than the
Diastolic Max
value.
Systolic Min
Lowest acceptable value for a systolic reading. The value entered must be
≥
2,
and must be greater than the
Diastolic Min
value.
Diastolic Max
Highest acceptable value for a diastolic reading. The value entered cannot
exceed 200 and must be less than
Systolic Max
value.
Diastolic Min
Lowest acceptable value for a diastolic reading. The value entered must be
≥
1,
and must be less than the
Systolic Min
value.
MAP Max
Highest acceptable value for a mean arterial pressure (MAP) reading. The value
entered cannot exceed 250.
MAP Min
Lowest acceptable value for a MAP reading. The value entered must be
≥
1.
PP Max
Highest acceptable value for a pulse pressure (PP) reading. The value entered
cannot exceed 250.
PP Min
Lowest acceptable value for a PP reading. The value entered must be
≥
1.
Heart Rate Max
Highest acceptable value for a heart rate reading. The value entered cannot
exceed 200.
Heart Rate Min
Lowest acceptable value for a heart rate reading. The value entered must be
≥
1.