MiPNet22.11
O2k-FluoRespirometer manual: supplement
S2
Oroboros Instruments
Mitochondria and Cell Research
Supplement B. O2k-Chamber assembly
Disassembly
1.
Remove the POS connectors from the POS holders by
pressing the sleeve downwards against the bottom plate
of the POS connector. Unplug the cable.
2.
Unscrew the blue
chamber holders
and the
POS holder
from the O2k housing.
3.
Carefully take out the glass chamber from the copper
block of the
O2k-FluoRespirometer
.
Assembly
1.
Insert the
glass chamber
into the copper block of the
O2k-FluoRespirometer
.
2
.
Screw the
POS holder
1 into a position where the glass
chamber with the angular cut towards the POS holder is
slightly lifted upwards.
3
. Screw the
chamber holder
loosely to the copper
block (arrow 2). Loosen the POS holder stepwise
(counterclockwise, not more than successive quarter
turns; arrow 1) and follow any downwards movement
of the glass chamber by screwing down the chamber
holder (arrow 2). When a slight backwards movement
of the POS holder 1 cannot be followed by any further
downwards movement of the chamber holder 2, then
screw the POS holder 1 clockwise (tight), thus securing
the glass chamber in a fixed position.
Clip A1
2
1