MiPNet22.11
O2k-FluoRespirometer manual
13
Oroboros Instruments
High-Resolution FluoRespirometry
O2k configuration window. The background is green if
rotation of the stirrer is on, briefly yellow when the stirrer
is accelerating, red if the stirrer is switched off.
O2 calib.
Yellow indicates that the POS has not been calibrated in
the present record. Calibration values are used as default
from files previously saved real-time.
L
Left double-click on O2 calib. to open the calibration
window (left or right).
O2 calib.
Green indicates that the POS has been calibrated within
the file. It can be re-calibrated any time when connected
or disconnected to the O2k.
Set temp.
37.0 indicates the temperature setting sent from the O2k
control window, green if the measured temperature is
within ±0.002 °C of the set temperature, but yellow if the
difference is more than ±0.002 °C.
Illum. on / Illum. off indicates whether the chamber illumination (not to
be confused with light sources for fluorescence) is on
(Illum.on, highlighted in yellow) or off (Illum.off, grey).
Toggle by a L
double-click. When the windows to the
glass chambers are closed with optical sensors, the
illumination is not visible and must be switched off.
O2k: Pan auto indicates that automatic panning is on (green) with the
O2k connected to DatLab. The time range is maintained
while the time axis always shows the currently recorded
data. The value of the offset (minimum time) increases
stepwise as experimental time proceeds.
O2k: Pan off: Yellow indicates that automatic panning is off. Plots are
shown in static display. This allows for manually panning
backwards to observe previous sections of the experiment
at a given time range. In this mode, the actual
experimental time may be off-scale. Toggle between Pan
auto
and
Pan off by a L
click on the text. This does not
influence continuous data recording.
TIP: Ready
Indicates that a Titration-Injection microPump (TIP2k) is
connected to the O2k, and is ready to start F8. For TIP2k
instructions see
MiPNet12.10
.
ERROR
Blinks red if an error has occurred. In this case, L
click
on ERROR to open the Error message window. In this
window, L
Close to remove the blinking error signal.