10
6.6
Direct setting of the output voltage value between 0 and 14 V, 0-560mV, 0-28mV and T/C
Sliding switch set to OUT-V.
The key Num activates the numeric selection.
Enter the required value with the keyboard and press Enter.
To erase the entry, use the keys
←
and
→
.
6.7
Selection of the display function
The key MENU selects following operation modes – see page 14:
TARA SELECT
Tara of the Multimeter input signal to Zero
FILTER DEGREE
Selection of averaging filter from 0 to 9.
RAMP SELECT
Activates Ramps:
1
Raising Ramp with 0.5mA, 0.25V, 10mV, 0.5mV and 20 ºC steps
2
Raising Ramp with 0.1mA, 0.05V, 2mV, 0.1mV and 5 ºC steps
3
Sinking Ramp with -0.1mA, -0.05V, -2mV, -0.1mV and -5 ºC steps
4
Sinking Ramp with -0.5mA, -0.25V, -10mV, -0.5mV and -20 ºC steps
GRAPHICS
Graphics of last 128 measuring points in the Multimeter function of the voltage
range up to 200V DC or currents up to 100mA DC.
BACKLIGHT
Back ground light: OFF or ON. With ON selection the display remains illuminated
for 20 seconds after pressing any key of the keypad.
VOLTAGE OUTPUT U
Output Signal selection
1
Output 0…14V
2
Output 0…560mV
3
Output 0…28mV
4
Thermocouples Simulation
THERMOCOUPLE OUTPUT
(Select 4 is activated)
1
Type B
2
Type E
3
Type J
4
Type K
5
Type N
6
Type R
7
Type S
8
Type T
Cold Junct. Comp.
Cold Junction Compensation can be activated or switched-off.
Unit Selection
ºC or ºF
EEprom Archive
Erase or Store the settings in internal EEPROM.
The key Esc terminates the programming mode and the display switches into the Measuring mode.