Calibrator
Current Source
0 ... 22 mA
Current Sink
0 ... 22 mA
Voltage Source
0 ... 14 V
mV Source
0 … 28.000 mV and 0 … 560.00 mV
DIN Thermometers J, K, N, R, S, T, B, E
The calibrator functions are settable with three sliding switches:
ON-OFF
Power ON and OFF
SRC-I, SNK-I
Current Source 0-22mA, Current Sink 0-22mA
OUT-V, OUT-I
Voltage Output 0-14V, Current Output 0-22mA
Multimeter
Voltage Ranges
Firm ranges ± 2V, ± 20V, ± 200VDC
Autoranging 0 ... ± 200VDC
Current Range
Firm range
0 ... ± 100mADC
The Multimeter Functions can be set with the keys 100mA, AUTO, 2V, 20V, 200V.
2
OUTPUTS and INPUTS
CALIBRATOR:
Voltage and Current Output: + OUT -
MULTIMETER:
Voltage Input:
GND, IN-V
Current Input:
GND, IN-I
3
LCD DISPLAY
The LCD Display is divided into 2 sections. The upper part shows the multimeter functions, the lower
part shows the generated calibrator signals. A Bargraph in the middle displays the measured values as
an analogue function. The left side of the display shows various statements:
AUTO-FIX
Automatic Range selection or Firm Ranges
200V
Firm Ranges 2V, 20V, 200V, 100mA
INP
The display shows the input signal. With FN the following functions can be set:
HLD
Display HOLD
MAX Maximum Value Hold
MIN
Minimum Value Hold
RST
Reset the max. and the min. values to zero
NUM
Direct value selection with the keyboard
RMP
Automatic Ramp
SUP
Charger plugged, Battery is charged
99%
Battery voltage shown in % of full capacity. Charge the battery when less
than 0% is displayed. Charge only with the enclosed original charger.
ATTENTION!
In the Current Sink Mode the switches SNK-I and OUT-I have to be selected. Do not switch to OUT-V!
The Current and the Voltage Outputs might be destroyed!