CALIBRATOR - FUNCTION SELECTION  

 
The  calibrator  functions  are  selectable  with  three  sliding  switches.  The  keyboard  permits  setting  of 
Ramps, Steps or direct Value entry. 
The Multimeter is active all the time and can be simultaneously used with the calibrator. The function is 
described in § 7. 
 
The  display  is  divided  into  two  parts.  The  upper  part  shows  the  multimeter  values,  the  lower  part  is 
assigned to the calibrator outputs 0-20mA or 0-10V. The middle part is a Bargraph showing the values 
as analogue information. 
 
6.1 

Setting of the output current 0 - 22mA in Steps or Ramps 

Sliding switch to be set to SRC-I and OUT-I. 
The key  Clr  set the output current to 4mA.  
The key  Up  increases the output current in 0.20 mA steps. 
The key  Down  decreases the output current in 0.20 mA steps. 
The key  Ramp  activates the automatic ramp. To stop the ramp, press the key again. 
The key  Step  generates the current steps of 4, 8, 12, 16, 20, 22mA 
 
6.2 

Direct setting of the output current between 0 and 22mA 

Sliding switch to be set to SRC-I and OUT-I. 
The key Num activates the numeric selection.  
Enter the required value with the keyboard and press  Enter
To erase the last entry, use the keys  

 and 

 . 

 
6.3 

Setting of the voltage output 0 - 14 V in Steps or Ramps 

Sliding switch to be set to OUT-V. 
The key  Clr  sets the output voltage to 0 V.  
The key  Up  increases the output voltage in 0.10 V steps. 
The key  Down  decreases the output voltage in 0.10 V steps. 
The key  Ramp  activates the automatic voltage ramp. 
The key  Step  generates voltage increments 0, 2, 4, 6, 8, 10, 12, 14V. 
 
6.4 

Setting of mV Outputs 0–560mV and 0–28mV in Steps or Ramps 

Same like 014V voltage output. Menu selection: Select 2 or Select 3. 
 
6.5 

Thermocouple output in Steps or Ramps 

Put the Sliding Switch into Position OUT-V. 
Press MENU until the display shows VOLTAGE OUTPUT U. 
Select 4 with MENU. 
THERMOCOUPLE OUTPUT. Select the required Type. 
Select Cold Junct. Comp. Select 1 (ON) for internal on-board compensation. Press MENU. 
Select Unit Selection. Select 1 for degree Centigrade (ºC). Press ESC.   
Key Clr sets the output signal to 0.0 ºC.  
Key Up increases the Output mV in 10 ºC steps. 
Key Down decays the Output mV in 10 ºC steps. 
Key Ramp activates the automatic Ramp generation. 

 

Key Step generates Output mV in Steps 0, 200, 400, 600, 800, 1000, 1200, 1400 ºC  
 
 
 
 
 
 
 
 

RAMPS 
The key Ramp activates the Ramp Generator. The display increments or decrement in steps selected 
in  the  Menu  and  shows  the  symbol  RMP.  With  the  key  Ramp  the  Ramp  can  be  stopped.  When 
pressed again, the Ramp will continue from the value when it had been stopped. By keeping pressed 
the key Ramp for several seconds, the Ramp will start from beginning.  

 

Summary of Contents for OC502-t

Page 1: ...2 t_GBM_21311 HANDHELD CALIBRATOR MULTIMETER OC502 t Owner s Manual ORBIT CONTROLS AG Zürcherstrasse 137 CH 8952 Schlieren ZH Tel 41 44 730 2753 Fax 41 44 730 2783 info orbitcontrols ch www orbitcontrols ch ...

Page 2: ...2 ...

Page 3: ... with banana and crocodile 300 mm length Owner s manual with Calibration Certificate NOTE When you receive the shipment inspect the container and equipment for signs of damage Note any evidence of rough handling in transit Immediately report any damage to the Orbit Controls customer service Phone 4144 730 2753 or Fax 4144 730 2783 and to the shipping agent The carrier will not honour damage claims...

Page 4: ...ps 9 6 6 Direct setting of the output voltage value 0 14 V 0 560mV 0 28mV and T C 10 6 7 Selection of the display function 10 7 MULTIMETER FUNCTION SELECTION 11 8 SOFTWARE CALIBRATION 11 8 1 Calibration of Calibrator 12 8 1 1 Current output 4 20mA 12 8 1 2 Current Sink 0 4 20m 12 8 1 3 Voltage Ranges 0 14V 0 560mV 0 28mV 13 8 2 Calibration of Multimeter Inputs 13 8 2 1 Current Input 0 100mA 13 8 2...

Page 5: ...th high resolution are available Din Thermocouples J K N R S T B E can be simulated in positive temperature range The sensor type will be selected and the required temperature entered with the keyboard The display shows the temperature the mV output is at the output terminals The junction is compensated for the ambient temperature with internal sensor The compensation can be switched off OC502 t c...

Page 6: ...e upper part shows the multimeter functions the lower part shows the generated calibrator signals A Bargraph in the middle displays the measured values as an analogue function The left side of the display shows various statements AUTO FIX Automatic Range selection or Firm Ranges 200V Firm Ranges 2V 20V 200V 100mA INP The display shows the input signal With FN the following functions can be set HLD...

Page 7: ...0 3 2 5 ºC Load maximum load 1kOhm Could Junction compensation with SMT160 can be menu calibrated Working Ranges and specified Accuracies for Thermocouples without internal compensation The accuracies shown are maximum values Typical accuracies are lower R Range C 50 0 0 400 400 1760 Accuracy C 1 6 1 6 1 6 S Range C 50 0 0 500 500 1760 Accuracy C 2 4 1 9 1 5 B Range C 50 800 800 100 1000 1820 Accu...

Page 8: ...ed with sliding switches The settings and the parameters are stored in internal memory The instrument can be software calibrated via the keyboard Keyboard The keys have to be pressed at least for 0 5 sec in order to permit the controller to accept the command Rate 1 sample per 1 sec Memory EEPROM SUPPLY Battery Two 3 7V 2000mAh rechargeable Li ion Batteries permit an uninterrupted operation of 9 h...

Page 9: ...in Steps or Ramps Sliding switch to be set to OUT V The key Clr sets the output voltage to 0 V The key Up increases the output voltage in 0 10 V steps The key Down decreases the output voltage in 0 10 V steps The key Ramp activates the automatic voltage ramp The key Step generates voltage increments 0 2 4 6 8 10 12 14V 6 4 Setting of mV Outputs 0 560mV and 0 28mV in Steps or Ramps Same like 0 14V ...

Page 10: ... 2mV 0 1mV and 5 ºC steps 4 Sinking Ramp with 0 5mA 0 25V 10mV 0 5mV and 20 ºC steps GRAPHICS Graphics of last 128 measuring points in the Multimeter function of the voltage range up to 200V DC or currents up to 100mA DC BACKLIGHT Back ground light OFF or ON With ON selection the display remains illuminated for 20 seconds after pressing any key of the keypad VOLTAGE OUTPUT U Output Signal selectio...

Page 11: ... maximum value MIN The display shows the minimum value RST The maximum and the minimum stored values are erased 8 SOFTWARE CALIBRATION The voltage and the current ranges of the multimeter and the output signals of the calibrator can be precisely calibrated by using the keyboard The calibration menu can be opened with the key Set A full calibration of all 5 ranges or selected ranges only can be app...

Page 12: ... display to calibration menu 8 1 2 Current Sink 0 4 20mA Sliding Switches OUT I and SNK I Output Terminal OUT connect with the Plus Terminal of external voltage source max 24V Output Terminal OUT connect with the Minus Terminal of external voltage source with an mA Meter connected in serial Press the key 1 The display shows With the keys Up or Down appears the calibration constant 0 To change it e...

Page 13: ...shows Short the inputs Press Enter to store the zero input Press Menu The display changes to Apply 100mA from external calibrator Press Enter to store the 100mA full range input 8 2 2 Voltage Input 2VDC Connect 2VDC from external Voltage Calibrator to Input Terminals GND and IN V Press 3 The display shows Short the inputs Press Enter to store the zero input Press Menu The display changes to Apply ...

Page 14: ... 12 bit or 16 bit signal resolution 20mA or 100mA output current and the possible correction of the cold junction compensation The Option 100mA has to be ordered from the manufacturer The factory setting is 16 bit When 12 bit resolution is selected the display and the output signal resolution is limited to two decimal points NOTE Do not approach the calibration when the battery sign 0 at the displ...

Page 15: ...ated The filter function is activated after the key Menu is pressed twice The display shows Key 0 switches the filter off Keys 1 9 select the filter value The filter selection is closed with Esc 13 RAMPS The Ramp function can be opened with Menu The display shows RAMP SELECT 1 Increasing ramp with steps of 0 5mA 0 25V 10mV 0 5mV and 20 ºC 2 Increasing ramp with steps of 0 1mA 0 05V 2mV 0 1mV and 5...

Page 16: ... supplied from OC502 V2 100 Both the generated calibration signal and the measured external signal are shown at the LCD digital display OC502 V2 100 can be software calibrated via the keyboard whenever required The instrument is supplied from internal battery and permits an uninterrupted operation of 3 4 hours with the output current between 0 and 20mA For higher output currents up to 100mA the ca...

Page 17: ... mA 8 000 mA 20 000 mA 20 000 mA VOLTAGE OUTPUT VOLTAGE OUTPUT VOLTAGE OUTPUT Display Output V Display Output mV Display Output mV 0 000 V 0 00 mV 0 000 mV 2 000 V 100 00 mV 5 000 mV 10 000 V 500 00 mV 25 000 mV THERMOCOUPLES B E J K N R S T 300ºC mV 0 ºC mV 1000 ºC MULTIMETER Max Inaccuracy 0 1 from Value 1 digit VOLTAGE INPUT CURRENT INPUT 2V DC 20V DC 200V DC 100 mA DC Input Display V Input Dis...

Page 18: ...will be pleased to answer all your questions not only to this instrument but also to all our calibrators and measuring equipment Please call our customer service or write to us info orbitcontrols ch Thank you PLEASE NOTE The instrument is supplied from internal rechargeable Li ion battery Please make sure that the battery is correctly charged from the enclosed battery charger as soon as the 0 sign...

Reviews: