13
8.1.3 Voltage Ranges 0-14V, 0-560mV, 0-28mV
Output Terminals + OUT – with external V-Meter
connected.
Sliding Switch in Position OUT-V.
Set the output voltage 2V with Up or Down
and store with Enter.
Select 10V Range with Menu, set the calibration
Constant with Up or Down and store with Enter.
Continue with Menu for calibration of Ranges
560mV and 28mV.
With the key Esc changes the display into the
calibration Menu.
8.2 CALIBRATION of MULTIMETER INPUTS
8.2.1 Current Input 0-100mA
Connect external Current Calibrator to Input Terminals GND and IN-I.
Press 2. The display shows:
Short the inputs. Press Enter to store the zero input.
Press Menu. The display changes to:
Apply 100mA from external calibrator.
Press Enter to store the 100mA full range input.
8.2.2 Voltage Input 2VDC
Connect 2VDC from external Voltage Calibrator to Input Terminals GND and IN-V.
Press 3. The display shows:
Short the inputs. Press Enter to store the zero input.
Press Menu. The display changes to:
Apply 2VDC from external calibrator.
Press Enter to store the 2V full range input.
8.2.3 Voltage Input 20VDC
Inputs GND and IN-V supplied from external voltage calibrator 20VDC.
Press 4. The display shows:
Short the inputs. Press Enter to store the zero input.
Press Menu. The display changes to:
Apply 20VDC from external calibrator.
Press Enter to store the 20V full range input.