20 

 

The brightness is not 
enough

 

 

The brightness is not adjusted 
accurately

 

Adjust the brightness adjustment 
knob

 

The position of condenser is too low

 

Adjust its position

 

The condenser is not centered

 

Adjust it to center

 

                                         

 

                                                   

A12.1095

 

6.2 

Mechanical Part

 

TROUBLE 

CAUSE 

SOLUTION 

The image can’t focus 
using high-magnification 
objective 

The slide is on the reverse side 

Reverse the slide 

The cover glass is too thick 

Use standard cover glass with 
thickness of 0.17mm 

The objective touch the 
cover glass when it 
change from low 
magnification to 
high-magnification 

The slide is on the reverse side 

Reverse the slide 

The cover glass is too thick 

Use standard cover glass with 
thickness of 0.17mm 

The binocular images do 
not coincide 

The interpupillar distance is 
not correct 

Adjust it correctly 

Eyes too tired 

The diopter is not proper 

Adjust the diopter properly 

The brightness of the 
illumination is not proper for 
eyes 

Adjust the lamp voltage 

The objective 
magnification does not 
match the screen display 
magnification 

The objective is not properly 
installed to the specified 
converter hole or the screen 
setting is inconsistent with the 
actual situation 

The objective is re installed to the 
corresponding hole position or 
reset on the screen to match the 
actual situation 

  

6.3 

Electric Part

 

TROUBLE 

CAUSE 

SOLUTION 

The LED lamp does not 
light when the switch is on

 

No power or the fuse burn out 

Check the connection of the   
power cord or the fuse 

The LED lamp is broken 

Replace it 

The brightness is not 
enough 

LED lamp attenuates after long 
time use 

Replace the LED lamp 

The bulb flickers   

The light bulb is going to wear 
out 

Replace it 

The LED lamp does not 
light when the switch is on 

No power or the fuse burn out 

Check the connection of the   
power cord or the fuse 

Can not take photo 

Is correctly installed correctly 

Install the corresponding drive 
according to the camera model 

Improper line connection 

Check if the link between the 
microscope, the computer and 
the camera is normal and 
reconnected 

The screen is always being 
loaded after turning on 

Improper stage connection 

Check whether the connection 
between the platform and the 
host is normal and reseat the 
connection well 

Summary of Contents for A12.1095

Page 1: ...ed Instruction Manual This manual is for users of A12 1095 scientific research microscope To ensure your safety obtain optimum performance and to familiarize yourself fully with the use of this micros...

Page 2: ...ing the extensional function board 11 3 11 Computer control signal interface and photo function key 11 3 12 Additional function keys 12 3 13 Touch screen 12 4 Installation 13 4 1 Installing Diagram 13...

Page 3: ...nvection air which is used for cooling 6 Make sure the instrument is earthed to avoid lightning strike 7 Check the input voltage be sure the input voltage which signed in the back of the microscope is...

Page 4: ...or potential sources of electrical sparks and use them in a drafty room as possible as you can 5 Don t use organic solvent to wipe the non optical elements when you need to clean use the soft detergen...

Page 5: ...4 1 Components Name A12 1095 Objective Eyepiece Viewing Head Condenser Nosepiec e Main Body Stage...

Page 6: ...peed control button 5 Condenser adjustment knob 6 Switch button for upper and lower light 7 Brightness adjustment knob 8 Condenser 1 2 3 4 5 6 7 8 9 10 11 12 13 9 LCD touch screen 10 Aperture iris dia...

Page 7: ...r condenser 3 Condenser adjustment knob 4 Adjustment hand wheel 5 Photo button 6 Filed iris diaphragm adjustment ring 7 Shortcut 8 Tension adjustment ring for lamp housing 9 Ground lever 10 Lamp housi...

Page 8: ...t on the top is opening the symbol is When the light switch button is push out the light on the bottom is opening the symbol is 3 3 Light socket Fig 3 As shown in Fig 3 there are red dots in the socke...

Page 9: ...e the objective may collide with the specimen 3 5 Adjusting the interpupillary distance Fig 5 The interpupillary distance range 47mm 78mm While looking through the eyepieces adjust for binocular visio...

Page 10: ...e converted to other magnification objective to achieve focusing without damaging the specimen 2 When camera is needed Push the light path switch selector rod as shown in Fig 7 to the end for binocula...

Page 11: ...he field iris diaphragm inscribed with the field of viewing That means the condenser is centred correctly 7 Slightly increase the field iris diaphragm in reality operation to make the image circumscri...

Page 12: ...out the standard bright and dark field board then inset the extensional function board into the slot and tighten the screw 3 11 Computer control signal interface and photo function key Fig 12 13 Conn...

Page 13: ...croscopes has a set of supporting software For the specific operation of the software please refer to the software manual The function keys can be used in combination with the touch screen Please refe...

Page 14: ...installation sequence of the components The number in the figure shows the assembly steps Before installing be sure every components is clean do not score any parts or glass surface Keep well with hex...

Page 15: ...l groove of the stage holder with the dovetail bulge of the frame and slide down naturally until lower limit screw is limited Then the screw I in Fig 2 is screwed inward with a wrench 4 2 2 Installing...

Page 16: ...photo attachment or C mount 0 5 Mount the photo attachment or 0 5 C mount according to the assembly path into the connecting seat of the trinocular viewing head and tighten the screw V During the inst...

Page 17: ...en instruction manual for specific operation Clean the objectives regularly for lens is susceptible to dust When operating use 10 magnification objective to search and focus specimen firstly then repl...

Page 18: ...erface into the ternary camera barrel to an appropriate position Then screw in the number XI bolt to fasten the camera interface In order to ensure the appropriate position of camera interface the spe...

Page 19: ...ctric stage 270 210mm Moving range 125 75mm Illumination 3W LED lamp Condenser Abbe condenser NA0 9 0 1 Software Image software Scopeimage10 Photo attachment Video attachment C Mount 1 C Mount 0 5 Pol...

Page 20: ...the reverse side Turn it over Dry objective has oil on it especially for 40X objectives Wipe the oil A lens the objective condenser eyepiece or collector is dirty Clean it Immersion oil is not used wi...

Page 21: ...en display magnification The objective is not properly installed to the specified converter hole or the screen setting is inconsistent with the actual situation The objective is re installed to the co...

Page 22: ...the instructions A12 1095 Stage is not working properly Improper line connection Check the internal connections and reconnect well Initialization cannot stop after turning on Improper control board R...