Performing Manual Measurements
The following section covers the minimum basics required to perform measurements. Additional features built
into the Lasercheck Windows software are discussed in detail further in the Lasercheck Windows software
manual.
Choose a Setup File
From the Main Screen use the “>” button to select the proper setup file. The controller is delivered with 4
factory provided setup files loaded – grind_st.set, uncal_r.set, uncal_s.set, and serial.set. For a detailed
discussion of creation and installation of setup files please see the section labeled “SETUP FILES” and the
“Appendix - Calibration Procedures” later in this manual.
This shows the GRIND_ST.SET file as the active setup file.
Measure Window with Results
Pushing the Meas button will perform a measurement and the following screen is displayed:
The Ra roughness of the surface that the measurement head is placed on is displayed in microinches (µin) or
microns (um) of roughness.
Saving a Measurement
Pushing the Select button saves last measurement in the control box. The following screen will be presented
after the file is saved:
Pressing the Meas button without first pushing the Select button will perform another measurement without
saving the last measurement in the control box.
SELECT:>DEFAULT
^System
Ra: 123.4
in
SAVE ^Exit
Saved filename
MEAS ^Exit