Onset
Computer
Corporation
User’s Guide
- 42 -
HOBO U30 Station/RC
A d d i n g a S e n s o r
New sensors will only be recognized by the HOBO U30 Station when it is not logging. If you attempt to
add a sensor while the logger is recording data, it will be ignored. Logging will continue normally for all
other sensors already plugged in.
Smart Sensors
To add new Smart Sensors:
1.
Using HOBOware Pro, stop the logger from logging
2.
Readout the data from the logger.
IMPORTANT
: All existing data will be lost when you re-launch the logger.
3.
Plug in any new Smart Sensors.
4.
Re-launch the logger. The new Smart Sensors should start recording measurements. Analog Sensors
To add new sensors to the
Analog Sensors
1.
Connect the HOBO U30 Station to a computer via the USB cable.
2.
In HOBOware Pro, select Stop Logging.
3.
Readout the data from the logger.
IMPORTANT
: All existing data will be lost when you re-launch the logger.
4.
Connect the new sensors to the Analog Sensor Port.
5.
Configure Analog Sensor Port for the new sensor.
6.
Re-launch the logger.
The new sensors should start recording measurements.
Summary of Contents for hobo u30