OMS LINEA PATAVIUM S.T.01/3X ED.0 REV.8 09/2017
66
IEC 61000-4-11
40 % U
T
(60 % hole in U
T
per 5 cicli)
70 % U
T
(30 % hole in U
T
per 20 cicli)
<5 % U
T
(>95 % hole in U
T
per 5S)
40 % U
T
5 cycle (100mS)
70 % U
T
20 cycle (500mS)
<5 % U
T
5S
magnetic field at Network
frequency (50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
The magnetic fields at network frequency should have
characteristic levels of a typical locality in a environment
such as commercial or hospital.
NOTE: U
T
is the network tention in c.a. before the application of test level.
Manufacturer's Guide and declaration – Electromagnetic immunity
The device is intended to work in the electromagnetic environment below specified. The customer or the user of the device must ensure that it is used in
this environment.
Immunity test
Level test IEC 60601
Level of
Compliance
Electromagnetic Environment-Guide
RF conduct
IEC 61000-4-6
RF irradiata
IEC 61000-4-3
3 Veff
from 150 kHz to 80 MHz
3 V/m
from 80 MHz to 2,5 GHz
3 V
3 V/m
Portable and mobile RF communication equipment should not
be used near any part of the device including cables, the
recommended separation distance calculated with the equation
applicable to the frequency of the transmitter.
Recommended
Distance Separation:
d = 1,17 √P
d = 1,17 √P from 80 MHz to 800 MHz
d = 2,34 √P from 800 MHz to 2,5 GHz
where P is the maximum rated output power of the transmitter
in Watts (W) according to the manufacturer of the transmitter
and D is the recommended separation distance, in meters (m).
The field intensities of fixed RF transmitters, as determined by
an electromagnetic investigation on site A should be below the
level of compliance for each frequency range B.
Interference may occur near devices marked with the following
symbol: