OMS LINEA PATAVIUM S.T.01/3X ED.0 REV.8 09/2017
44
ADJUSTMENTS
ADJUSTMENTS
Activity
Frequency
By whom?
Note/references
General surgery water tap
End of the work day Trained
operator
Close the dental unit water supply
tap
Instrument spray water pressure
adjustment
When necessary
Trained
operator
Tap under the instrument table
(adjustment without tool)
Instrument air pressure adjustment
When necessary
Authorised
technician
Tap under the instrument table
(adjustment with tool, authorised
technician only)
Clutch adjustment (accessible from the
exterior)
When necessary
Trained
operator
Cup/spittoon water flow adjustment
When necessary
Trained
operator
Oscillating arm spring adjustment
(interior)
When necessary
Authorised
technician
Main unit water pressure adjustment
When necessary
Authorised
technician
Pressure regulator calibration
Main unit air pressure adjustment
When necessary
Authorised
technician
Pressure regulator calibration
INSTRUMENT SPRAY PRESSURE
The water delivered to the spray is regulated by rotating the tap located under the relevant instrument
module: when turned anticlockwise, the flow of water to the spray is gradually reduced; when turned
clockwise, the opposite applies.