OMS LINEA PATAVIUM S.T.01/3X ED.0 REV.8 09/2017
53
INSTRUMENT WATER CONTAINER DISINFECTION
Note:
We do not recommend you keep distilled water in the container for prolonged periods of
time. This could generate a dangerous bacterial load.
Note:
Frequently check and sanitise the container; if deteriorated or deformed, replace it with an
original OMS container.
MAIN UNIT WATER FILTER
Note:
Small impurities may be present in the water mains and may stop in the water filter inside the
main unit. This causes clogging and, consequently, a reduction of the available water flow. If
this is the case, clean the filter located inside the chromium-plated fitting at the base of the
quadruple solenoid valve in the main unit. This operation must be performed by a skilled
technician during periodic equipment overhaul.