Through-beam Sensor Heads: FUN Mode Functions
Section 3-8
83
Correcting Display Values to Match Actual Width: Two-point Scaling A
Display values can be corrected if there is a discrepancy between the
actual incident beam width and the value displayed on the Amplifier Unit.
When actual widths are known, they are input at two points to correct the
range and offset of display values (see following figure).
To change only the offset without changing the range of
display values, refer to
Offsetting Display Values: One-
Example: Two-point Scaling A
Sensor Head Emitter
Sensor Head Emitter
Sensor Head Receiver
Sensor Head Receiver
Sensing object 1
Despite the actual 2.0-mm incident
beam width, 2.2 mm is displayed.
Sensing object 2
In addition, despite the actual 4.0-mm
incident beam width, 3.6 mm is displayed.
: Scaling value inputs
2.0
2.5
3.0
3.5
4.0
2.0
2.5
3.0
3.5
4.0
2.0
2.5
3.0
3.5
4.0
2.20
2.55
2.90
3.25
3.60
2.20
2.55
2.90
3.25
3.60
Actual incident beam width
Actual incident beam width
Measured value = Display value
Display value
Measured value
There are discrepencies
between actual incident beam
width and display values.