![background image](http://html1.mh-extra.com/html/omron/r8it/r8it_instruction-manual_741896025.webp)
25
Troubleshooting
Trouble
Possible cause
How to correct
Replace the batteries with
Batteries are worn out.
new ones.
Nothing is displayed
(See to Page 6.)
when you push the
START/STOP botton
The directions of
Insert the batteries correctly.
polarities of inserted
(Refer to Page 6.)
batteries (+ and -) are wrong.
The wrist cuff is not held
Measure your blood pressure
at the level of the heart.
again in correct posture.
(Refer to Page 9.)
Measurement cannot be
made, or the blood pressure
The wrist cuff is not
Wrap the wrist cuff correctly.
values are displayed
wrapped snugly.
(Refer to Page 8.)
abnormally high (low).
Too much stress is applied
Relax yourself and make
on your shoulder or arm.
measurement.
You are talking or moving
Be still and make
your hands during
measurement.
measurement.
The blood pressure
values differ from
Refer to the section on "Wrist Blood Pressure Monitor."
the values meas
(Refer to Page 28.)
ured at brachial.
Blood pressure values
are displayed before
Air is leaking from the
Contact the store where the
the monitor inflates
wrist cuff.
unit was sold or the nearest
to the required value.
OMRON dealer.
Blood pressure value
Blood pressure value tends to fluctuate depending on the
fluctuates each time of
mental state during measurement or the measuring time.
measurement. Or
Breathe deeply to relax yourself and make measurement
the blood pressure
again.
values are displayed
(Refer to Page 27.)
abnormally high (low).
• If you cannot resolve the problem even after taking the above-mentioned measures, please
contact the nearest OMRON dealer.
Summary of Contents for R8IT
Page 29: ... R8IT OMRON ...
Page 30: ...2 3 4 5 6 7 8 9 10 13 14 16 20 23 23 24 25 26 27 ...
Page 32: ...4 01234 5 01234 90 01 23456789 USB 0 1 2 ...
Page 33: ...5 USB R8IT AAA ...
Page 34: ...6 USB 1 2 2 AAA LR03 3 012 345 0123456 ...
Page 35: ...7 1 START STOP 2 2001 2030 3 SET 012 4 5 SET 01 6 7 SET 8 START STOP ...
Page 36: ...8 0123 0 123 18 012 1 0 012 10 15 mm 2 012345 6 145 mm 012 0 012345 3 01 2 18 1 2 3 10 15mm ...
Page 37: ...9 0 0 ...
Page 38: ...10 012345 1 START STOP 7 2 ...
Page 39: ...11 0123 4 3 9 ...
Page 40: ...12 4 START STOP START STOP ...
Page 41: ...13 90 MEMORY 91 0123 0 19 OFF 17 012 345 1 MEMORY 2 GRAPH MEMORY ...
Page 44: ...16 OMRON R8IT 0123 SET SET 0 123 START STOP SET START STOP 17 18 18 18 19 19 19 ...
Page 45: ...17 ON OFF R8IT 0123456789 1 SET 2 ON OFF OFF OFF ON 3 SET SET START STOP ...
Page 52: ...24 012 10 10 10 6 OMRON E 01234567 OMRON ...
Page 53: ...25 6 6 9 8 28 OMRON 27 START STOP 01234567 OMRON ...