Omnitronix DL50 User’s Manual
Page 6
Data-Link DL50 Data Alarm Field Definition
Data Field: A
A) Start Position [0]
B) Field Length [0]
C) Field Name []
3. Select Start Position. When prompted to enter a new value, enter “1” and press <ENTER>.
4. Select Field Length. When prompted to enter a new value, enter “4” and press <ENTER>.
5. Select Alarm Name and enter “test_field”, then press <ENTER>.
6. Press <ENTER> to return to the Field definition Table. If configured properly, the data alarm
field should appear in this menu.
7. Press <ENTER> to return to the Data Alarm Settings menu. From here, select Alarm 1. The
following menu will be displayed:
Data-Link DL50 - Data Alarm 1
A) Alarm Enable [OFF]
B) Alarm Name []
C) Alarm Equation []
D) Threshold [1]
E) Auto-Clear when Threshold Reached [ON]
F) Alarm Counter Clear Interval [12 HOURS]
G) Alarm Counter Reset Time [00:00]
H) Actions []
I) Message []
J) Clear This Alarm Counter Now
8. Select Alarm Enable to toggle the value to ON.
9. Select Alarm Name and enter “Data Alarm 1”.
10. Select Actions and enter “T1” to cause this data alarm to send a trap to SNMP manager #1, as
configured in the SNMP managers menu.
11. Select Alarm Equation and enter “
test_field=’test’
”. This will cause an alarm to occur
any time the word “test” is received.
12. All other settings should remain at their default values for this test scenario.
13. Finally, the serial ports must be configured to monitor for data alarms. Return to the root of
the setup menu and select “Serial Settings” then “Serial Port 1 Settings”.
14. Toggle “Data Alarm Enable” to the value of 1. This causes the DL50 to monitor data
received on port 1 for alarms in data alarm family 1.
15. Press <CTRL> + C to return to the command processor.
Testing
Connect to the unit via serial connection on I/O 1 and enter the string “test” and then <ENTER>.
This should trigger the above data alarm, and an SNMP trap should be sent to trap manager #1. If
this is not the case, repeat the Network and Data Alarms settings.
Note:
There will be a 30 second delay in alarming if the terminal emulator being used does not send a
LF with the CR. This may be circumvented by pressing <CTRL + J>.