1.3.3

Power supply

Operating the 

CPC

 without PE and ground connection is life-threatening and not permitted.

► Only operate the 

CPC

 with a mains power supply connected to protective earth (PE).

Power supply from grounded grids (TN/TT)

Before a measurement is started, the 

CPC

 automatically verifies the PE connection in grounded grids 

(TN/TT). 
► If this check fails, check the power cord and power supply.

If the error message persists, there is no intact connection to protective earth (PE). 
This is life-threatening. In this case measurements are not permitted and cannot be performed.

Power supply from isolated grids (IT)

An IT grid is a grid structure where none of the active conductors are galvanically connected to ground. 
In an IT grid, only the PE is connected to ground.

In IT grids, the check fails – even if there is a PE connection. This can be the case when the 

CPC

 is 

powered by a generator. Since every operation mandates a PE connection for the operation of the 

CPC

you need to manually verify this.

If the 

CPC

 is supplied by a generator, the equipotential ground or PE of the generator has to be grounded 

properly. 
► If this is not possible, measurements are not permitted and cannot be performed.

Additional information

Instead of supplying the 

CPC

 from phase-neutral (L1-N, A-N), it may also be supplied from phase-phase 

(for example, L1-L2; A-B). 
► Make sure that the voltage does not exceed 240V AC.
► Make sure that the power supply is fuse-protected (16 A automatic circuit breaker).
► Do not use an extension cable on a cable reel to prevent an overheating of the cord; run out the 

extension cord.

► Keep extension cables as short as possible to prevent power loss.

If the power supply is 

 190V AC,

 

the

 CPC

 cannot provide the full output power at the 

800A AC

 output. 

The same applies when an external current booster is used. 
► Therefore, in order to gain the full output power, provide a sufficient power supply 

(190 V ... 240 V AC).

The 

Ext. Booster 

connector is 

always

 galvanically connected to mains and active. 

This also applies when no external booster is selected, the green status light (0) is on, the outputs are 
turned off, or the Emergency Stop button is pressed.

► Handle the 

Ext. Booster

 connector with extreme caution.

► Only use booster cables supplied by OMICRON.
► Do not use booster cables that are frayed or damaged in any way.

Summary of Contents for CPC 100

Page 1: ...CPC Getting Started...

Page 2: ...ta embodied in this manual represent the technical status at the time of writing and are subject to change without prior notice We have done our best to ensure that the information given in this manua...

Page 3: ...d the regulations on site 1 2 Safety standards and rules 1 2 1 Safety standards Testing with the CPC must comply with the internal safety instructions and additional safety relevant documents In addit...

Page 4: ...yee of the utility Turn off and disconnect the high voltage from the test object Protect yourself and your working environment against an accidental re connection of high voltage by other persons and...

Page 5: ...o the grounding system on site This also applies to all other test devices and accessories in the test setup Make sure that the grounding clamp has a good electrical contact to the grounding system on...

Page 6: ...the equipotential ground or PE of the generator has to be grounded properly If this is not possible measurements are not permitted and cannot be performed Additional information Instead of supplying t...

Page 7: ...e to be connected to the CPC 100 do not carry any voltage potential During a test the only power source for a test Make sure that the ground terminal of the device under test is in good condition clea...

Page 8: ...t amount of heat approx 300 W m at 800 A After operation wait for cables and clamps to cool down before touching them If in doubt wear gloves Never use the test cards Quick and Resistance to measure t...

Page 9: ...nues the monitoring of the discharge status before the test object might be completely discharged The CPC displays an error message and a green light in this case This could erroneously be interpreted...

Page 10: ...struction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be require...

Page 11: ...100 hardware concept In contrast to the CPC 100 the CPC 80 only comes with the booster output It is not equipped with any measurement inputs or directly usable outputs and therefore intended for use w...

Page 12: ...anel control Fuses for 6A AC and 6A DC outputs Navigation elements Test Start Stop Soft touch keyboard Key lock BIN IN Binary trigger input Emergency stop button I O Status lights INPUT Measuring inpu...

Page 13: ...Figure 3 4 Left side view of the CPC 80 ePC with front panel control Key lock I O Status lights Soft touch keyboard Navigation elements Test Start Stop Emergency stop button External booster Groundin...

Page 14: ...mains Table 3 1 Status lights 1 Output is switched OFF Green status light 0 on Current voltage source is inactive off 2 Output is switched ON and or measurement process is active Red status light I is...

Page 15: ...r is always galvanically connected to mains This also applies when no external booster is selected the green status light 0 is on the outputs are turned off or the Emergency Stop button is pressed Han...

Page 16: ...00 and CPC 80 USB connector Connector for OMICRON safety accessories RJ 45 socket for connection to a PC or a network hub Yellow LED Data are being transferred from to the network Green LED CPC connec...

Page 17: ...test object according to your requirements 5 Set up a barrier to separate the high voltage area from the work area see Figure 1 1 Separation of work area and high voltage test area on page 9 6 Remove...

Page 18: ...for the first time a warning message appears on the screen Only after the I O test start stop push button is pressed a second time the voltage is applied to the CPC output An error message 313 appears...

Page 19: ...the CPC and the test object 5 Remove the barrier 4 2 Cleaning Use a cloth dampened with isopropanol alcohol to clean CPC and its accessories WARNING Death or severe injury caused by high voltage or c...

Page 20: ...CPC Getting Started 20 OMICRON...

Page 21: ...3 59495 4444 Additionally you can find our Service Center or Sales Partner closest to you at www omicronenergy com Customer Portal stay informed www omicronenergy com customer The Customer Portal on o...

Page 22: ...ENU 1194 03 04...

Reviews: