CPC Getting Started
14
OMICRON
3.5
Status lights
► Always observe the status lights while working with the
CPC
.
► Do not cover the status lights during operation.
If no or both status lights are on, the unit is defective or not supplied by mains.
Table 3-1: Status lights
1. Output is switched OFF
Green status light (0)
on
Current/voltage source is
inactive/off
2. Output is switched ON and/or
measurement process is active
Red status light (I) is
flashing
Dangerous
operating condition
If the
CPC 100
is supplied by mains and switched on, and
no
or
both
status lights are on, the unit might be defective.
► Contact OMICRON support (see "Support" on page 21).
Dangerous
operating condition