Safety instructions

1.4

Operating the measurement setup

► If the CPC 100 or any add-on device or accessory does not seem to function properly, do not use it 

anymore. Please call the OMICRON hotline.

► If you have a cardiac pacemaker, do not use the CPC! If you have another type of electronic medical 

implant, consult your doctor before operating the CPC. Make sure there is no person with an 
electronic medical implant such as a cardiac pacemakerin the immediate vicinity.

► Before handling the 

CPC

 in any way, connect its grounding terminal to the grounding system on site.

► Make sure that the grounding clamp has a good electrical contact to the grounding system on site 

and avoid connecting it to corroded or painted surfaces.

► Make sure that a test object’s terminals that are to be connected to the CPC 100 do not carry any 

voltage potential. During a test, the only power source for a test 

► Make sure that the ground terminal of the device under test is in good condition, clean and free of 

oxidation.

► Protect others from accessing the high-voltage test area and accidentally touching live parts by 

setting up a suitable barrier and, if applicable, warning lights. 

► Do not enter the high-voltage test area if the red status light of the 

CPC

 is on since all outputs carry 

dangerous voltage or current.

Figure 1-1: Separation of work area and high-voltage test area

► Do not open the 

CPC

 housing.

► Do not repair, modify, extend, or adapt the 

CPC

 or its accessories.

► Use only the original 

CPC

 accessories available from OMICRON.

► Use the 

CPC

 and its accessories only in a technically sound condition. 

Make sure its use is in accordance with the regulations on site and the designated use described in 
this document.

CPC

Device under test

Work area

High-voltage test area

Summary of Contents for CPC 100

Page 1: ...CPC Getting Started...

Page 2: ...ta embodied in this manual represent the technical status at the time of writing and are subject to change without prior notice We have done our best to ensure that the information given in this manua...

Page 3: ...d the regulations on site 1 2 Safety standards and rules 1 2 1 Safety standards Testing with the CPC must comply with the internal safety instructions and additional safety relevant documents In addit...

Page 4: ...yee of the utility Turn off and disconnect the high voltage from the test object Protect yourself and your working environment against an accidental re connection of high voltage by other persons and...

Page 5: ...o the grounding system on site This also applies to all other test devices and accessories in the test setup Make sure that the grounding clamp has a good electrical contact to the grounding system on...

Page 6: ...the equipotential ground or PE of the generator has to be grounded properly If this is not possible measurements are not permitted and cannot be performed Additional information Instead of supplying t...

Page 7: ...e to be connected to the CPC 100 do not carry any voltage potential During a test the only power source for a test Make sure that the ground terminal of the device under test is in good condition clea...

Page 8: ...t amount of heat approx 300 W m at 800 A After operation wait for cables and clamps to cool down before touching them If in doubt wear gloves Never use the test cards Quick and Resistance to measure t...

Page 9: ...nues the monitoring of the discharge status before the test object might be completely discharged The CPC displays an error message and a green light in this case This could erroneously be interpreted...

Page 10: ...struction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be require...

Page 11: ...100 hardware concept In contrast to the CPC 100 the CPC 80 only comes with the booster output It is not equipped with any measurement inputs or directly usable outputs and therefore intended for use w...

Page 12: ...anel control Fuses for 6A AC and 6A DC outputs Navigation elements Test Start Stop Soft touch keyboard Key lock BIN IN Binary trigger input Emergency stop button I O Status lights INPUT Measuring inpu...

Page 13: ...Figure 3 4 Left side view of the CPC 80 ePC with front panel control Key lock I O Status lights Soft touch keyboard Navigation elements Test Start Stop Emergency stop button External booster Groundin...

Page 14: ...mains Table 3 1 Status lights 1 Output is switched OFF Green status light 0 on Current voltage source is inactive off 2 Output is switched ON and or measurement process is active Red status light I is...

Page 15: ...r is always galvanically connected to mains This also applies when no external booster is selected the green status light 0 is on the outputs are turned off or the Emergency Stop button is pressed Han...

Page 16: ...00 and CPC 80 USB connector Connector for OMICRON safety accessories RJ 45 socket for connection to a PC or a network hub Yellow LED Data are being transferred from to the network Green LED CPC connec...

Page 17: ...test object according to your requirements 5 Set up a barrier to separate the high voltage area from the work area see Figure 1 1 Separation of work area and high voltage test area on page 9 6 Remove...

Page 18: ...for the first time a warning message appears on the screen Only after the I O test start stop push button is pressed a second time the voltage is applied to the CPC output An error message 313 appears...

Page 19: ...the CPC and the test object 5 Remove the barrier 4 2 Cleaning Use a cloth dampened with isopropanol alcohol to clean CPC and its accessories WARNING Death or severe injury caused by high voltage or c...

Page 20: ...CPC Getting Started 20 OMICRON...

Page 21: ...3 59495 4444 Additionally you can find our Service Center or Sales Partner closest to you at www omicronenergy com Customer Portal stay informed www omicronenergy com customer The Customer Portal on o...

Page 22: ...ENU 1194 03 04...

Reviews: