2.3
Connectors on the front side
WARNING
Death or severe injury caused by high voltages possible.
All inputs and outputs of
TICRO 100
are electrically connected to the SELV (safety
extra low voltage) insulation group of the device.
►
Do not connect any voltages that are not SELV compliant.
Connector
Description
10 MHz output (50
Ω
BNC socket).
This output provides a highly accuracte 10 MHz sine wave signal
(4 dBm ± 2 dB at 50
Ω
). The 10 MHz output is always active and cannot be
configured in the Web Interface.
Output 1 and 2 (50
Ω
BNC socket).
Outputs 1 and 2 can be configured in the Web Interface for the following
signals:
- IRIG-B, unmodulated TTL signal or modulated on 1 kHz carrier
- PPX (TTL signal, adjustable number of pulses, for example, 1 PPS or 1 kHz)
- DCF77 unmodulated TTL signal
- Trigger (TTL signal, 1 pulse at a defined point in time)
- Trigger (TTL signal, PPX sequence starting at the defined point in time)
For information about configuring these outputs, please refer to section
"Output configuration page" of the TICRO 100 User Manual.
Output 3 (optocoupler output 30 V/100 mA max., terminal block supplied on
delivery).
Output 3 can be configured in the Web Interface for the following signals:
- IRIG-B unmodulated
- PPX (adjustable number of pulses, for example, 1 PPS or 1 kHz)
- DCF77 unmodulated
- Trigger (1 pulse at a defined point in time)
- Trigger (PPX sequence starting at the defined point in time)
For information about configuring this output, please refer to section "Output
configuration page" in the TICRO 100 User Manual.
TICRO 100 Quick Start Guide
10
OMICRON