C
URSOR
M
EASUREMENTS
Two cursors can be used to make instantaneous Time and Voltage measurements
for CH1 and CH2 signals. Refer to the following table for more information.
Function menu
Settings Description
Type OFF
Voltage
Time
Exit the cursor measurement mode
Display voltage measurement cursor and menu
Display time measurement cursor and menu
Source CH1,
CH2
Select the waveform channel on which the cursor
measurement will be performed
Delta (MS460)
Display the difference between CH1 and CH2 values
Cursor 1
(MS460)
Data for cursor 1
Cursor 2
(MS460)
Data for cursor 2
To use the cursor for a
Voltage
measurement on CH1, perform the following steps:
1. Press
the
MENU
key; the function menu will then display at the right of the
screen.
2. Press
the
MENU
▲
or
MENU
▼
key to select
Curs Meas
. Five (5) selectable
options (MS46) or two (2) selectable options (MS420) appear at the bottom of
the screen.
3. Press
the
F1
key to select the measurement type
Voltage
. Two dashed lines
V1
and
V2
will appear on the screen.
4. Press
the
F2
key to select
CH 1
.
5.
Press OPTION and the display will show the following:
Cursor 2
Cursor 1
Use
▲
(yellow) and
▼
(yellow) to move V1 up and down; the voltage value
(relative to the zero position) will display at the bottom of the screen. Use
◄
(yellow) and
►
(yellow) to move V2 up and down; the voltage value (relative
to the zero position) will display at the bottom of the screen. The display will
also show the result of V1 - V2. Refer to Figure 36-a.
Summary of Contents for HOS-MS420
Page 11: ...10 V5 1 March 2010 FRONT PANEL AND KEYPAD OVERVIEW Figure 3 ...
Page 32: ...31 V5 1 March 2010 Figure 20 AC Coupling ...
Page 72: ...71 V5 1 March 2010 Fig 42 Rectangle window Fig 43 Blackman window Fig 44 Hanning window ...
Page 79: ...78 V5 1 March 2010 ...
Page 85: ...84 V5 1 March 2010 Figure 4 Continuity Test ...
Page 89: ...88 V5 1 March 2010 ...
Page 91: ...90 V5 1 March 2010 Figure 10 400mA AC Current Measurement ...
Page 95: ...94 V5 1 March 2010 ...