Refer to Figure 30:
Figure 30: Save and Recall Waveforms
F
UNCTION
S
ETTING
M
ENU
(
REFER
TO
T
ABLE
BELOW
)
Press the
MENU
button and use the up/down arrow keys to scroll to
FUNCTION
Function menu
Description
Recall Factory
Set the meter’s settings to their factory default
conditions
Auto Calibration Perform an Auto-calibration procedure
Language
Select the desired display language (English, Chinese,
Deutsch, or Greek)
Auto-calibration
The Auto-calibration function automatically configures internal parameters to maintain the
sensitivity and accuracy of the meter. Use the Auto-calibration in the following cases:
•
When the temperature changes by more than 5 degrees Celsius during operation
•
When operating the oscilloscope in a new bench top or field environment
1. Press
MENU
and select the
FUNCTION
menu using the
MENU
▲
or
MENU
▼
key
2. Press
F2
(Auto calibration). A message appears asking the user to remove all cables