6
Dip the
HHWT-486672
strip into the
CELL
and immediately press
READ
. This starts the
20
SECOND
countdown timer. During this time move the strip in a gentle back and forth motion
(approx. 2 strokes/Sec).
Remove and discard the strip after “1” on the display disappears*
.
The meter will begin counting up for
120 seconds
, at the end of which, the cursor will move across
the display while the meter prepares to measure the sample. Record result displayed (this result
is automatically stored in HR8). After testing, rinse cell immediately and clean with brush.
*NOTE:
8VHDVHFRQGGLSWLPHLIZDWHUWHPSHUDWXUHLVDERYHÛ&
DIP STRIP -
(read carefully and follow procedure closely)
5
Press the
ZERO/ON
button. The cursor will move across the display, followed by
0.00 PPM
.
The sample is ready for testing.
ZERO METER*
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the
MENU
button until the display shows the parameter HR8.
SELECT TEST: HR8
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\P/
with the water sample.
FILL METER WITH SAMPLE
11
High Range Chlorine Test Procedure
HR
HR8
1
Remove one (1)
HHWT-486672
strip from the bottle before beginning the test. Set the strip
in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
pH Test Procedure
PH
PH7
*
NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.
6
Dip the
HHWT-486639
into the
CELL
and immediately press
READ
. This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle back and forth motion (approx. 2
strokes/Sec).
Remove and discard the strip after “1” on the display disappears*
. The cursor
will move across the display while the meter prepares to measure the sample. Record result
displayed (this result is automatically stored in PH7 Menu). After testing, rinse cell immediately.
127()RUEHVWUHVXOWV7RWDO$ONDOLQLW\RIWKHVDPSOHVKRXOGEHSSP
*
NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.
DIP STRIP AND PRESS “READ”
5
Press the
ZERO/ON
button. When the display shows
0.0 PH
, the sample is ready for testing.
ZERO METER*
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\P/ZLWK
the water sample.
FILL METER WITH SAMPLE
3
Press and re-press the
MENU
button until the display shows the parameter
PH7
.
SELECT TEST: PH7
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
1
Remove one (1)
HHWT-486639
from the bottle before beginning the test. Set the strip in a dry,
convenient place and recap the bottle immediately.
REMOVE STRIP