DPD-1 Free Chlorine Test Procedure

NOTE:

  N,N-diethyl-p-phenylenediamine (DPD) also reacts to form a magenta color with Ozone, Total Chlorine, 

Permanganate, Iodine, and other oxidizers.

DO NOT

 discard the sample from the Free Chlorine test if you are planning to run 

HHWT-486638 (DPD-3) Total Chlorine Procedure. Move directly to steps 8-10 on page 7. 
Otherwise, rinse the cell immediately.

Remove one (1) 

HHWT-486637

 strip from the bottle before

beginning the test. Set the strip in a dry, convenient place
and recap the bottle immediately.

REMOVE STRIP

Press the 

ZERO/ON

 button to power the meter on; the display

will show all annunciators, then the current MENU selection,
followed by the last reading.

TURN METER ON

Dip the 

HHWT-486637 

strip into the CELL and  immediately

press READ. This starts the 

20 SECOND 

countdown timer.

During this time move the strip in a gentle back and forth motion
(approx. 2 strokes/Sec). 

Remove and discard the strip after

“1” on the display disappears.*

DIP STRIP AND PRESS “READ”

Press the 

ZERO/ON

 button. The cursor will move across the

display followed by 

0.00 PPM

. Sample is ready for testing.

ZERO METER*

Rinse the 

CELL

 at least 3 times with the water sample you will

be testing - rinsing minimizes the potential for cross-
contamination from a previous test. Finally, fill cell to capacity
(4mL) with the water sample.

RINSE AND FILL CELL WITH SAMPLE

Press and re-press the 

MENU

 button until the display shows

the parameter 

CL1

.

SELECT TEST: CL1

CL1

1

2

6

7

The cursor will move across the display while the meter
prepares to measure the sample. Record result displayed
(this result is automatically stored in CL1 Menu).

RECORD RESULT DISPLAYED

5

4

3

CL

*

NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.

Summary of Contents for HHWT-11

Page 1: ...HHWT 11 Handheld Chlorine Photometer e mail info omega com For latest product manuals www omegamanual info Shop online at omega comSM User sGuide...

Page 2: ...pecifications without notice omega com info omega com Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Cana...

Page 3: ...HP6 pH 11 PH7 High Range Chlorine HRC 11 HR8 6SHFL FDWLRQV About Your Photometer 5 Tips for Best Accuracy 12 About the Built In Cell 12 Meter Messages 13 Battery Installation 13 Reorder Information 13...

Page 4: ...GHG Battery Life WHVWV ZLWK DONDOLQH EDWWHULHV Electromagnetic Compliance PLWWHG QWHUIHUHQFH 1 EMC PPXQLW WR QWHUIHUHQFH 1 Waterproof Rating FHHGV 3 Weight QVWUXPHQW J R Dimensions QVWUXPHQW FP LQ HHW...

Page 5: ...O SDUDPHWHUV AWWA 4500 Cl G Strip DPD 1 Anhydrous DPD sulfate Anhydrous Na2HPO4 Anhydrous KH2PO4Na2 EDTA 1 5 33 4 64 0 1 1 1 5 33 4 64 0 1 1 Component Free Chlorine About The Accuracy Calibration Of T...

Page 6: ...20 SECOND countdown timer During this time move the strip in a gentle back and forth motion approx 2 strokes Sec Remove and discard the strip after 1 on the display disappears DIP STRIP AND PRESS READ...

Page 7: ...oxide CL1 Chlorine and Iodine react with N N diethyl p phenylenediamine as it is released from the strip to form a magenta color directly proportional to the Chlorine concentration Ozone Bromine and P...

Page 8: ...he 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion approx 2 strokes sec Remove and discard the strip after 1 on the display disappears The cursor will move...

Page 9: ...ON 1 Remove one 1 HHWT 486633 strip and one 1 HHWT 484014 strip from their bottles before beginning the test Set the strips in a dry convenient place and recap the bottles immediately REMOVE STRIPS C...

Page 10: ...SUHYLRXV WHVW LQDOO OO FHOO WR FDSDFLW P ZLWK the water sample FILL METER WITH SAMPLE HP HP6 PA PA5 6 Dip the HHWT 486675 strip into the CELL and immediately press READ This starts the 20 SECOND coun...

Page 11: ...ry convenient place and recap the bottle immediately REMOVE STRIP pH Test Procedure PH PH7 NOTE When testing outdoors sunlight for best accuracy use the Mixing Cap Cell Cover when Zeroing and Reading...

Page 12: ...EO VPDOOHU RU ODUJHU LQ width than the normal strips in the bottle These should be discarded Using these strips may give incorrect results The HHWT 11 Photometer is not compatible for use with DPD 1 D...

Page 13: ...igh quality AAA alkaline batteries be used 4 Replace the battery cover Be sure to tighten the cover securely Do not over tighten This is necessary for meter to be waterproof 5 Dispose of the used batt...

Page 14: ...00 5 00 6 00 0 00 1 00 2 00 3 00 4 00 5 00 6 00 Hach DR890 ppm HHWT 11 ppm mg L ppm mg L DR890 HHWT 11 0 00 0 00 0 41 0 39 0 92 0 96 0 79 0 73 1 28 1 32 2 70 2 73 3 20 3 22 4 45 4 75 5 30 5 53 Meter M...

Page 15: ...otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 16: ...ity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Reviews: