DIP STRIP AND PRESS “READ”

About 4 seconds after the meter auto-zeroes, “20” will appear on the display. Immediately dip the 

HHWT-486633 

strip into the 

CELL

. During this 

20 SECOND

 countdown timer,

 move the strip in 

a gentle back and forth motion (approx 2 strokes/sec). 

Remove and discard the strip after “1” 

on the display disappears*

The meter will begin counting up for 100 seconds, at the end of 

which

 the cursor will move across the display while the meter prepares to measure the sample. 

Record result displayed (this result is automatically stored in Cd4). After testing, rinse cell 
immediately and clean with brush.

5

Dip the 

HHWT-484014

 Glycine strip into the 

CELL

 and immediately press 

READ

. This starts the

20 SECOND

 countdown timer. During this time move the strip in a gentle back and forth motion

(approx 2 strokes/sec). 

Remove and discard the strip after “1” on the display disappears

*.

The meter will auto-zero. Get ready to dip 

HHWT-486633

 strip.

DIP STRIP AND PRESS “READ”

4

Rinse the 

CELL

 at least 3 times with the water sample you will be testing - rinsing

PLQLPL]HVWKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWR

capacity (4mL) with the water sample.

FILL METER WITH SAMPLE

3

Press and re-press the 

MENU

 button until the display shows the parameter 

Cd4

.

SELECT TEST: Cd4

2

Press the 

ZERO/ON

 button to power the meter on; the display will show all annunciators,

then the current MENU selection, followed by the last reading.

TURN METER ON

1

Remove one (1) 

HHWT-486633

 strip and one (1) 

HHWT-484014

 strip from their bottles before

beginning the test. Set the strips in a dry, convenient place and recap the bottles immediately.

REMOVE STRIPS

Chlorine Dioxide Test Procedure

1

Remove one (1) 

HHWT-486634

 strip from the bottle before beginning the test. Set the strip

in a dry, convenient place and recap the bottle immediately.

REMOVE STRIP

6

DIP STRIP AND PRESS “READ”

Dip the 

HHWT-486634

 strip into the 

CELL

 and immediately press 

READ

. This starts the 

20 

SECOND

 countdown timer. During this time move the strip in a gentle back and forth motion 

(approx 2 strokes/sec). 

Remove and discard the strip after “1” on the display disappears*

The cursor will move across the display while the meter prepares to measure the sample. 
Record result displayed (this result is automatically stored in O3). After testing, rinse cell 
immediately and clean with brush.

5

Press the 

ZERO/ON

 button. The cursor will move across the display, followed by 

0.00 PPM

.

The sample is ready for testing.

ZERO METER*

2

Press the 

ZERO/ON

 button to power the meter on; the display will show all annunciators, then

the current MENU selection, followed by the last reading.

TURN METER ON

3

Press and re-press the 

MENU

 button until the display shows the parameter 

O3

.

SELECT TEST: O3

4

Rinse the 

CELL

 at least 3 times with the water sample you will be testing - rinsing minimizes

WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\P/ZLWK

the water sample.

FILL METER WITH SAMPLE

Ozone Test Procedure

o

3

O3

*

NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.

Summary of Contents for HHWT-11

Page 1: ...HHWT 11 Handheld Chlorine Photometer e mail info omega com For latest product manuals www omegamanual info Shop online at omega comSM User sGuide...

Page 2: ...pecifications without notice omega com info omega com Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Cana...

Page 3: ...HP6 pH 11 PH7 High Range Chlorine HRC 11 HR8 6SHFL FDWLRQV About Your Photometer 5 Tips for Best Accuracy 12 About the Built In Cell 12 Meter Messages 13 Battery Installation 13 Reorder Information 13...

Page 4: ...GHG Battery Life WHVWV ZLWK DONDOLQH EDWWHULHV Electromagnetic Compliance PLWWHG QWHUIHUHQFH 1 EMC PPXQLW WR QWHUIHUHQFH 1 Waterproof Rating FHHGV 3 Weight QVWUXPHQW J R Dimensions QVWUXPHQW FP LQ HHW...

Page 5: ...O SDUDPHWHUV AWWA 4500 Cl G Strip DPD 1 Anhydrous DPD sulfate Anhydrous Na2HPO4 Anhydrous KH2PO4Na2 EDTA 1 5 33 4 64 0 1 1 1 5 33 4 64 0 1 1 Component Free Chlorine About The Accuracy Calibration Of T...

Page 6: ...20 SECOND countdown timer During this time move the strip in a gentle back and forth motion approx 2 strokes Sec Remove and discard the strip after 1 on the display disappears DIP STRIP AND PRESS READ...

Page 7: ...oxide CL1 Chlorine and Iodine react with N N diethyl p phenylenediamine as it is released from the strip to form a magenta color directly proportional to the Chlorine concentration Ozone Bromine and P...

Page 8: ...he 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion approx 2 strokes sec Remove and discard the strip after 1 on the display disappears The cursor will move...

Page 9: ...ON 1 Remove one 1 HHWT 486633 strip and one 1 HHWT 484014 strip from their bottles before beginning the test Set the strips in a dry convenient place and recap the bottles immediately REMOVE STRIPS C...

Page 10: ...SUHYLRXV WHVW LQDOO OO FHOO WR FDSDFLW P ZLWK the water sample FILL METER WITH SAMPLE HP HP6 PA PA5 6 Dip the HHWT 486675 strip into the CELL and immediately press READ This starts the 20 SECOND coun...

Page 11: ...ry convenient place and recap the bottle immediately REMOVE STRIP pH Test Procedure PH PH7 NOTE When testing outdoors sunlight for best accuracy use the Mixing Cap Cell Cover when Zeroing and Reading...

Page 12: ...EO VPDOOHU RU ODUJHU LQ width than the normal strips in the bottle These should be discarded Using these strips may give incorrect results The HHWT 11 Photometer is not compatible for use with DPD 1 D...

Page 13: ...igh quality AAA alkaline batteries be used 4 Replace the battery cover Be sure to tighten the cover securely Do not over tighten This is necessary for meter to be waterproof 5 Dispose of the used batt...

Page 14: ...00 5 00 6 00 0 00 1 00 2 00 3 00 4 00 5 00 6 00 Hach DR890 ppm HHWT 11 ppm mg L ppm mg L DR890 HHWT 11 0 00 0 00 0 41 0 39 0 92 0 96 0 79 0 73 1 28 1 32 2 70 2 73 3 20 3 22 4 45 4 75 5 30 5 53 Meter M...

Page 15: ...otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 16: ...ity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Reviews: