HHTQ35 Digital Torque Gauges
User’s Guide
16
6. The display will appear as follows:
Optionally exercise the sensor several times (at full scale, if possible), then press
ENTER
.
7. The display will appear as follows:
Apply torque equal to the full scale of the instrument, then press
ENTER
.
8. After displaying “PLEASE WAIT…” the display will appear as follows:
Remove the torque applied in Step 7, leave the fixtures in place, then press
ZERO
.
9. The display will appear as follows:
Use the
UP
and
DOWN
keys to adjust the torque value as required. The torque values default to
even increments, as indicated by the previously entered number of data points described in Step
1. Then press
ENTER
.
Repeat the above step for the number of data points selected.
10. After all the clockwise calibration points have been completed, the display will appear as follows:
CALIBRATION
CLOCKWISE
GAIN ADJUST
APPLY FULL SCALE LOAD
50.000 LBFIN +/-20%
THEN PRESS ENTER
CALIBRATION
CLOCKWISE
ENSURE NO LOAD
THEN PRESS ZERO
CALIBRATION
CLOCKWISE
Optionally exercise
load cell a few times.
THEN PRESS ENTER
CALIBRATION
CLOCKWISE
APPLY LOAD
1 OF 5
ENTER LOAD:
10.000 LBFIN
THEN PRESS ENTER