HHTQ35 Digital Torque Gauges
User’s Guide
15
Selecting “CANCEL” will revert back to the Calibration setup. Selecting “EXIT W/O SAVING” will
return to the menu without saving changes.
3. After the number of calibration points has been entered, press
ENTER
. The display will appear as
follows:
4. Place the torque sensor vertically in a fixture free from vibration, then press
ZERO
. The gauge
will calculate offsets, and the display will appear as follows:
If failed:
5. The following screen appears after the offsets have been calculated:
Keep the sensor in a vertical position, as explained in Section 10.1. Attach weight fixtures
(brackets, hooks, etc), as required. Do not yet attach any weights or apply any calibration loads.
Then press
ENTER
.
CALIBRATION
NOT COMPLETE
CANCEL
EXIT W/O SAVING
CALIBRATION
CLOCKWISE
Attach necessary
weight fixtures.
THEN PRESS ENTER
CALIBRATION
OFFSET
Sen.Offset Adj.Failed
Ana.Offset Adj.Failed
CALIBRATION
OFFSET
Sen.Offset Adj.Passed
Ana.Offset Adj.Passed
CALIBRATION
OFFSET
Please wait…
CALIBRATION
OFFSET
Place gauge
vertical
THEN PRESS ZERO