background image

PRINCIPLE OF OPERATION

The transducer is a force balance device in which a coil is suspended in the field of a magnet by a flexure. Current flowing
through the coil generates axial movement of the coil and flexure. The flexure moves against the end of a nozzle, and creates
a back pressure in the nozzle by restricting air flow. This back pressure acts as a pilot pressure to an integral booster relay.

Consequently, as the input signal increases (or decreases, for reverse acting), output pressure increases proportionally.

Zero and span are calibrated by turning easily accessible adjusting screws on the front face of the unit. The zero adjustment
causes the nozzle to move relative to the flexure. The span adjustment is a potentiometer that limits the flow of current
through the coil. A thermistor circuit in series with the coil provides temperature compensation.

MOUNTING

The transducer can be mounted at any angle, but should be
calibrated after mounting. For maximum output pressure
stability, it should be mounted vertically in a vibration free

location or such that the vibration is isolated to the X and Z
axis as shown in 

Figure 3

.

Pipe:

Due to its light weight , the transducer may be sup-

ported by the piping used for the supply and output.

Panel:

(with access to rear of panel) Attach transducer

(

See Fig. 3

), to panel using two 10-32 screws and two

threaded mounting holes on the back of the unit.

Panel:

(with no access to rear of panel) Attach bracket

(

see Fig 1 and 3

), to transducer using two 10-32 screws

on the back of the unit and mount bracket to panel using
four 8-32 screws.

AIR CONNECTIONS

Supply Air must be instrument quality air regulated
between the minumum and maximum specified on the
nameplate of the transducer.

Instrument quality air consists of :

a  A dew point less than 35˚F
b  No particles larger than three microns
c  Maximum oil content of 1 ppm

All unused ports must be plugged.

Supply:

Connect supply to the1/4" NPT ports marked “IN” on the base of the transducer. Avoid getting pipe sealant inside

the piping or transducer.

Output:

Connect output to the 1/4" NPT ports marked “OUT” on the base of the transducer. The pipe plugs may be removed

from the two auxiliary output ports for a pressure gauge or as an alternative output port.

M4110/

 IP211, EP211 SERIES

0519

A

3

Figure 1

Summary of Contents for EP211 Series

Page 1: ...e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide IP211 EP211 SERIES ELECTROPNEUMATIC CONVERTERS...

Page 2: ...h America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 3...

Page 3: ...free location or such that the vibration is isolated to the X and Z axis as shown in Figure 3 Pipe Due to its light weight the transducer may be sup ported by the piping used for the supply and output...

Page 4: ...ntBody EN 50081 1 1992 Test results confirmed that no precautions need to be observed during installation regarding electro magneticemissionsfromtheI P E P211Transducers EN 50082 1 1992 The following...

Page 5: ...erative that the output pressure be closely monitored as turning the zero screw excessive ly could damage the flexure 2 Apply the 100 signal to unit and adjust Span adjusting screw until output is at...

Page 6: ...0 25 C COIL 300 J2 J1 I P SCHEMATIC RT1 R1 68 FIGURE 2 BLACK WHITE RV1 121 R1 100 25 C COIL 300 J2 J1 E P SCHEMATIC The input impedance will vary depending upon the potentiometer setting of the SPAN R...

Page 7: ...put pressure increases then follow calibration procedure 2 Electrical connection has opened or wrong type of signal applied 2 Check electrical connections to insure that there is a correct input signa...

Page 8: ...M4110 0519 IP211 EP211 SERIES A 8 Figure 3...

Page 9: ...oisture or vibration improper specification misapplication misuse or other operating conditions outside of OMEGA s control Components in which wear is not warranted include but are not limited to cont...

Page 10: ...Air Velocity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps I...

Reviews: