background image

 

 

CL3515R K/J/T/E/R/S/N/L/U/B/C Calibrator Thermometer 

7. “T1/T2 SOURCE” button 

Pressing T1/T2 SOURCE to cycle through T1, T2 and SOURCE. In the main display 
the blinking digit is the one to be adjusted, you can push the “

” button to make 

right or left shifts to the desired position. When incrementing to the utmost range of 
the selected thermocouples, the LIMIT will show on the display. 
SOURCE is to provide the output parameter settings. There are ten individual 
temperature setting points in group 0, which can be set at your desired output point. 
Use “

” to shift the desired digit to be adjusted and rotate the knob to increase or 

decrease the values you want to set. Press the T/C OFFSET to save the settings. 
 

8. “OFFSET” button (Thermocouple offset adjust) 

When the main display input is T1 or T2, and socket thermocouple is connected. 
Press T/C OFFSET over two seconds the SET annouciator will appear on the right 
side of display and enter the offset adjustment mode. And the blinking digit is the one 
to be adjusted. Rotate the knob to the right increasing the values, to the left side 
decreasing the values. The maximum range of the knob is ±5 centigrade. When 
turning to the utmost range, it will appear LIMIT symbol on the left side of the 
display and means that there is no further incrementing of the offset. Press the T/C 
OFFSET over two seconds to save the settings. 
 

9. “MIN/MAX” button 

Press MIN/MAX button to enter the MIN/MAX recording mode and REC shows on 
the display. The beeper emits a tone when a new minimum or maximum measurement 
is recorded. Press the MIN/MAX button again to rotate through the current readings:
MAX: The highest measurement recorded. 
MIN: The lowest measurement recorded. 
MAX-MIN: The difference of the highest and the lowest measurement. 
AVG: The average values of the measurements. 
Press MIN/MAX button over two seconds to exit the function. 
 

10. Knob usages in the settings   

In the TYPE mode, it is used for thermocouples selection to make right or left shifts 
to choose selectors. In the SOURCE mode, it is used to increase or decrease the 
values of the output function. 
 

11. PWM Group 

Group 0 set 

In the OUT mode, press T/C OFFSET over two seconds to set. 
LCD display Set CLEAr press T1/T2 SOURCE button to clear data, display SEt 0-0. 
In the main display the blinking digit is the one to be adjusted, you can push the 

” button to make right or left shifts to the desired position. Rotary knob to 

increase or decrease the values. Press T1/T2 SOURCE to save the one step setting. 
CL3515R can set 10 step, (Group 1 to 9 use software setting) press T/C OFFSET exit 
group set mode. 

 

 

CL3515R K/J/T/E/R/S/N/L/U/B/C Calibrator Thermometer 

IV Measurement 

 

1. T1, T2, Channel Subtract (NONE, T1-T2, T2-T1) 

ALARM function

 

Disable     
Hi enable

 

Enables alarm Hi function, and it will become red flashing, when it is over Hi 
setting value. 

Lo enable

 

Enables alarm Lo function, and it will become blue flashing, when it is below Lo 
setting value. 

Both enable

 

To enable both Hi and Lo, and it will become red flashing, when it is over Hi setting 
value, and it will become blue flashing when it is below Lo setting value.   

HOLD 

Hold the present reading. 

Clear 

Clear MAX, MIN, AVG recorded. 

11 

 

Summary of Contents for CL3515R

Page 1: ...CL3515R K J T E R S N L U B C Calibrator Thermometer e mail info omega com For latest product manuals www omegamanual info Shop online at omega comSM User sGuide ...

Page 2: ...USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide omega com info omega com The information contained in this document is believed to be correct but OMEGA accepts no liability for any errors it contains and reserves the right to alter specifications without notice ...

Page 3: ...ensor The thermometer features a dual thermocouple input an adjustable T C offset The thermocouples types comply with the N I S T Monograph 175 Revised to ITS 90 standard SAFETY INFORMATION It is recommended that you read the safety and operation instructions before using the thermometer WARNING To avoid electrical shock do not use this instrument when working voltages at the measurement surface o...

Page 4: ...Hz to1000MHz the measurement number producing the unsteady flutter it immediately recover after stop the test Accessories Two type K thermocouple bead wires Two type K thermocouple calibration bead wires Maximum insulation temperature 260 C 500 F Wire accuracy 2 2 C or 0 75 of reading whichever is greater from 0 C to 800 C A 9 volts battery An instruction manual Pc connection software Pc interface...

Page 5: ...5 rdg 2 0 F 58 F to 346 F N TYPE 0 05 rdg 1 0 C 50 C to 0 C 0 05 rdg 0 5 C 0 C to 1300 C 0 05 rdg 2 0 F 58 F to 32 F 0 05 rdg 1 0 F 32 F to 2372 F R S B C TYPE 0 05 rdg 2 C 0 C to 1767 C 0 05 rdg 4 F 32 F to 3212 F 3 K TYPE 0 1 200 C to 1372 C or 328 F to 2501 F J TYPE 0 1 210 C to 1200 C or 346 F to 2192 F T TYPE 0 1 250 C to 400 C or 418 F to 752 F E TYPE 0 1 250 C to 1000 C or 418 F to1832 F R ...

Page 6: ...ion 10uV Accuracy 0 025 1 digit Temperature Coefficient 0 1 times the applicable accuracy specification per C from 0 C to 18 and 28 C to 50 C 32 F to 64 F and 82 F to 122 F Top Side 1 Output 1 2 Output 2 3 Input T1 4 Input T2 5 RS232 optical interface OPERATING INSTRUCTIONS 4 4 2 1 3 5 CL3515R CL3515R K J T E R S N L U B C Calibrator Thermometer 4 Record Export Both Display drawing T1 and T2 T1 Di...

Page 7: ...L N S R E T J Press the type button again to choose the selected thermocouple 4 Backlight button Pressing the button less than two seconds to turn on and pressing the button again less than two seconds to turn off the backlight in the LCD It will turn off after thirty minutes without operation Auto power off It is in the APO mode when the meter is turned on and will turn off the meter without oper...

Page 8: ...its a tone when a new minimum or maximum measurement is recorded Press the MIN MAX button again to rotate through the current readings MAX The highest measurement recorded MIN The lowest measurement recorded MAX MIN The difference of the highest and the lowest measurement AVG The average values of the measurements Press MIN MAX button over two seconds to exit the function 10 Knob usages in the set...

Page 9: ...al shock disconnect the thermocouple connectors from the thermometer before removing the cover MAINTENANCE Battery Replacement Power is supplied by a 9 volt battery NEDA 1604 IEC 6F22 The appears on the LCD display when replacement is needed To replace the battery remove the two screws from the back of the meter and lift off the battery cover Remove the battery from battery contacts Cleaning Perio...

Page 10: ... L U B C Calibrator Thermometer I CL3515R Software Install Click Next Click Next 8 CL3515R K J T E R S N L U B C Calibrator Thermometer Click Next Click Finish II Uninstall StarÆsettingÆcontrolÆnew remove programÆCL3515R 9 ...

Page 11: ... L U B C Calibrator Thermometer I CL3515R Software Install Click Next Click Next 8 CL3515R K J T E R S N L U B C Calibrator Thermometer Click Next Click Finish II Uninstall StarÆsettingÆcontrolÆnew remove programÆCL3515R 9 ...

Page 12: ...al shock disconnect the thermocouple connectors from the thermometer before removing the cover MAINTENANCE Battery Replacement Power is supplied by a 9 volt battery NEDA 1604 IEC 6F22 The appears on the LCD display when replacement is needed To replace the battery remove the two screws from the back of the meter and lift off the battery cover Remove the battery from battery contacts Cleaning Perio...

Page 13: ...its a tone when a new minimum or maximum measurement is recorded Press the MIN MAX button again to rotate through the current readings MAX The highest measurement recorded MIN The lowest measurement recorded MAX MIN The difference of the highest and the lowest measurement AVG The average values of the measurements Press MIN MAX button over two seconds to exit the function 10 Knob usages in the set...

Page 14: ...L N S R E T J Press the type button again to choose the selected thermocouple 4 Backlight button Pressing the button less than two seconds to turn on and pressing the button again less than two seconds to turn off the backlight in the LCD It will turn off after thirty minutes without operation Auto power off It is in the APO mode when the meter is turned on and will turn off the meter without oper...

Page 15: ...ion 10uV Accuracy 0 025 1 digit Temperature Coefficient 0 1 times the applicable accuracy specification per C from 0 C to 18 and 28 C to 50 C 32 F to 64 F and 82 F to 122 F Top Side 1 Output 1 2 Output 2 3 Input T1 4 Input T2 5 RS232 optical interface OPERATING INSTRUCTIONS 4 4 2 1 3 5 CL3515R CL3515R K J T E R S N L U B C Calibrator Thermometer 4 Record Export Both Display drawing T1 and T2 T1 Di...

Page 16: ...r 418 F to 752 F E TYPE 0 1 250 C to 1000 C or 418 F to1832 F R TYPE 1 0 C to 1767 C or 32 F to 3212 F S TYPE 1 0 C to 1767 C or 32 F to 3212 F N TYPE 0 1 200 C to 1300 C or 328 F to 2372 F L TYPE 0 1 200 C to 900 C or 328 F to 1652 F U TYPE 0 1 200 C to 600 C or 328 F to 1112 F B TYPE 1 600 C to 1820 C or 1112 F to 3308 F C TYPE 1 0 C to 2316 C or 32 F to 4200 F Based on the ITS 90 temperature st...

Page 17: ...Hz to1000MHz the measurement number producing the unsteady flutter it immediately recover after stop the test Accessories Two type K thermocouple bead wires Two type K thermocouple calibration bead wires Maximum insulation temperature 260 C 500 F Wire accuracy 2 2 C or 0 75 of reading whichever is greater from 0 C to 800 C A 9 volts battery An instruction manual Pc connection software Pc interface...

Page 18: ...ensor The thermometer features a dual thermocouple input an adjustable T C offset The thermocouples types comply with the N I S T Monograph 175 Revised to ITS 90 standard SAFETY INFORMATION It is recommended that you read the safety and operation instructions before using the thermometer WARNING To avoid electrical shock do not use this instrument when working voltages at the measurement surface o...

Page 19: ...otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA is not intended to be used nor shall it be used 1 as a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications or used on humans...

Page 20: ...ewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial pH Conductivity Equipment DATA ACQUISITION M U Data Acquisition Engineering Software M U Communications Based Acquisition Systems M U Plug in Cards for Apple IBM Compatibles M U Data Logging Systems M U Reco...

Reviews: