CL3515R K/J/T/E/R/S/N/L/U/B/C Calibrator Thermometer
5. Set State
Read
Read out the groups reading of CL3515R.
Save
Make the setting value read in CL3515R.
EXIT
6. Meter
Temperature
Unit
°C, °F, °K
Type
K, J, T, E, R, S, N, L, U, B, C, mV
Group
Group Item
Group 0 to Group 9
Unit
°C, °F, °K
14
CL3515R K/J/T/E/R/S/N/L/U/B/C Calibrator Thermometer
SPECIFICATIONS
ELECTRICAL
Temperature Scale:
Celsius or Fahrenheit user-selectable.
Measurement Range:
Accuracy:
K/J/T/E/L/U-TYPE
±(0.05% rdg + 0.5°C) -50°C to 1372°C
±(0.05% rdg + 1.0°C) -50°C to -250°C
±(0.05% rdg + 1.0°F) -58°F to 2501°F
±(0.05% rdg + 2.0°F) -58°F to -346°F
N-TYPE
±(0.05% rdg + 1.0°C) -50°C to 0°C
±(0.05% rdg + 0.5°C) 0°C to 1300°C
±(0.05% rdg + 2.0°F) -58°F to 32°F
±(0.05% rdg + 1.0°F) 32°F to 2372°F
R/S/B/C-TYPE
±(0.05% rdg + 2°C) 0°C to 1767°C
±(0.05% rdg + 4°F) 32°F to 3212°F
3
K-TYPE(0.1°) -200°C to 1372°C or -328°F to 2501°F
J-TYPE(0.1°)
-210°C to 1200°C or -346°F to
2
192°F
T-TYPE(0.1°) -250°C to 400°C or -418°F to 752°F
E-TYPE(0.1°) -250°C to 1000°C or -418°F to1832°F
R-TYPE(1°)
0°C to 1767°C or 32°F to 3212°F
S-TYPE(1°)
0°C to 1767°C or 32°F to 3212°F
N-TYPE(0.1°) -200°C to 1300°C or -328°F to 2372°F
L-TYPE(0.1°) -200°C to 900°C or -328°F to 1652°F
U-TYPE(0.1°) -200°C to 600°C or -328°F to 1112°F
B-TYPE(1°)
600°C to 1820°C or 1112°F to 3308°F
C-TYPE(1°)
0°C to 2316°C or 32°F to 4200°F
Based on the ITS-90 temperature standard.