CL3515R K/J/T/E/R/S/N/L/U/B/C Calibrator Thermometer
7. “T1/T2 SOURCE” button
Pressing T1/T2 SOURCE to cycle through T1, T2 and SOURCE. In the main display
the blinking digit is the one to be adjusted, you can push the “
” button to make
right or left shifts to the desired position. When incrementing to the utmost range of
the selected thermocouples, the LIMIT will show on the display.
SOURCE is to provide the output parameter settings. There are ten individual
temperature setting points in group 0, which can be set at your desired output point.
Use “
” to shift the desired digit to be adjusted and rotate the knob to increase or
decrease the values you want to set. Press the T/C OFFSET to save the settings.
8. “OFFSET” button (Thermocouple offset adjust)
When the main display input is T1 or T2, and socket thermocouple is connected.
Press T/C OFFSET over two seconds the SET annouciator will appear on the right
side of display and enter the offset adjustment mode. And the blinking digit is the one
to be adjusted. Rotate the knob to the right increasing the values, to the left side
decreasing the values. The maximum range of the knob is ±5 centigrade. When
turning to the utmost range, it will appear LIMIT symbol on the left side of the
display and means that there is no further incrementing of the offset. Press the T/C
OFFSET over two seconds to save the settings.
9. “MIN/MAX” button
Press MIN/MAX button to enter the MIN/MAX recording mode and REC shows on
the display. The beeper emits a tone when a new minimum or maximum measurement
is recorded. Press the MIN/MAX button again to rotate through the current readings:
MAX: The highest measurement recorded.
MIN: The lowest measurement recorded.
MAX-MIN: The difference of the highest and the lowest measurement.
AVG: The average values of the measurements.
Press MIN/MAX button over two seconds to exit the function.
10. Knob usages in the settings
In the TYPE mode, it is used for thermocouples selection to make right or left shifts
to choose selectors. In the SOURCE mode, it is used to increase or decrease the
values of the output function.
11. PWM Group
Group 0 set
In the OUT mode, press T/C OFFSET over two seconds to set.
LCD display Set CLEAr press T1/T2 SOURCE button to clear data, display SEt 0-0.
In the main display the blinking digit is the one to be adjusted, you can push the
“
” button to make right or left shifts to the desired position. Rotary knob to
increase or decrease the values. Press T1/T2 SOURCE to save the one step setting.
CL3515R can set 10 step, (Group 1 to 9 use software setting) press T/C OFFSET exit
group set mode.
6
CL3515R K/J/T/E/R/S/N/L/U/B/C Calibrator Thermometer
IV Measurement
1. T1, T2, Channel Subtract (NONE, T1-T2, T2-T1)
ALARM function
Disable
Hi enable
Enables alarm Hi function, and it will become red flashing, when it is over Hi
setting value.
Lo enable
Enables alarm Lo function, and it will become blue flashing, when it is below Lo
setting value.
Both enable
To enable both Hi and Lo, and it will become red flashing, when it is over Hi setting
value, and it will become blue flashing when it is below Lo setting value.
HOLD
Hold the present reading.
Clear
Clear MAX, MIN, AVG recorded.