92
Problem
Cause
Remedy
Page
7. Stage
a) The observation image moves
too much when the stage is
touched with hand.
A stage clutch lever is released.
Set both clutches securely.
36
The modules (stage, etc.) are loosened.
Secure the modules firmly.
–
The microscope is used in an
environment with frequent vibration.
Install the microscope in an
environment with less vibration.
–
The microscope is installed on a table
with a thin top board.
Install the microscope on a table with
a thick top board.
–
b) The X- and/or Y-feed knobs
run idle.
A clutch lever is released.
Set the clutches securely.
36
A clutch is set to an intermediate position. Release the clutches and retry setting
them securely.
36
c) The stage moves
spontaneously when a clutch
is released.
The horizontality of the base is not
adjusted.
The horizontality of the base is not
adjusted.
62
Adjust the horizontality of the desk
where the frame is installed.
–
d) The specimen is moved when
a knob is turned quickly.
The specimen is moved at the stage
movement speed
Fix the specimen.
–
8. Display (Digital indicator, Basic software and Excel transfer tool)
a) The display shows nothing.
The main switch of the control box is not
set to ON.
Set the main switch to “ " (ON).
22
The fuses in the control box are blown.
Replace the fuses.
73
Cables are not connected completely.
Connect them properly.
69
b) The Excel transfer tool does
not start.
Cables are not connected completely.
Connect them properly.
69
The COM port is used for other functions.
Check the settings of the COM port
again.
–
c) The measured values are
too different from what is
expected.
The “μm/mm/inch/mil” setting is
erroneous.
Set it properly.
49
d) The values cannot be
transferred to the controller.
Microsoft Office is not started.
Start Microsoft Office.
–
The version other than Microsoft Office
2007, 2010, 2013, 2016, or 2019 is installed.
Install the version later than
Microsoft Office 2007.
e) Only the X- and Y-axis values
are output to the external
processing devices.
The data output is set to the X-/Y-axes
only.
Set it properly.
49
f) Display shows "013F0500".
The setting value is not inputted.
Input the setting value.
74
g) Display shows "013F1503".
The hand switch STM7-HS is not
connected to the control box for manual
STM7-CB.
Connect the cable.
69
h) Display shows "013F1504".
The focus controller STM7-MCZ is not
connected to the control box for motorized
STM7-CBA.
Connect the cable.
69
i) Display shows "013F1505".
The Z-scale cable is disconnected.
Connect the cable.
69
j) Display shows "013F1506".
The emergency stop switch is pressed.
Reset the emergency stop switch.
22
k) Display shows "013F1507".
Since the emergency stop switch was
pressed before exiting last time, the
confirmation of operations when the
microscope starts is interrupted.
Operate either the FOCUS button
of the focus controller, the AF
START/STOP button or the fine/coarse
focusing knob.
22