22
33
Operation procedures
3-1 Turning Power ON
1
Functioning of Modules
Confirm followings before turning ON the power.
·
Ensure that the connection cables and the power cords
of the control box and power supply unit are connected
properly.
·
When the origin search is set to ON, if the power of the
control box for motorized STM7-CBA is turned ON, be careful
the focusing unit moves upward. (Motorized frame only)
For the origin search setting, refer to page 51.
·
Rotate the emergency stop switch
a
on the control
box for motorized STM7-CBA in the resetting direction
(clockwise) to ensure that the switch is not set to ON.
(Motorized frame only)
If the emergency stop switch is ON, refer to page 7 and
restore it.
CAUTION
Set the main switch b on the rear of the control box STM7-CB/STM7-
CBA to " " (ON). (Only when using the control box.)
Set the main switches of the modules you are using to " " (ON).
(For details, refer to the instruction manual provided with each
module.)
Confirm that the LED illuminator is turned ON by using the light
intensity control function of the hand switch STM7-HS or the focus
controller STM7-MCZ.
For details, refer to "3-3 Hand switch (Manual frame only)" (page 25) or
"3-4 Focus controller (Motorized frame only)" (page 28).
The display c on the digital indicator should be illuminated now.
Rotate the X-axis knob d , the Y-axis knob e , the fine/coarse
focusing knob f or the fine/coarse focusing knob of the focus
controller STM7-MCZ to make sure that the display on the digital
indicator shows counting.
1
2
3
4
a
b
c
d
e
f
} The operation procedures are illustrated in combination with the middle manual frame as a reference. Unless otherwise