71
STM7
For product safety, connect the round-shape terminals of the functional ground cables to the
mark
area between the stage and the frame and between the frame and the control box, and tighten the
screws with the Allen screwdriver provided with the microscope frame.
(In case of STM7-CS50 or STM7-CS100, the connection destination of the functional ground cable is at
the rear of the stage.)
CAUTION
Bend the functional ground cables so that the excess
force is not applied to the foot of the round shape
terminal or the bonding area of the round shape terminal
and also they do not collide with other cables or cords.
CAUTION
When connecting interface cables of the microscope
digital camera DP22/DP27 to the controller, connect them
to the connector of USB 3.0.
When connecting USB cables of digital camera STM7-CU
to the controller, connect them to the connector of USB
2.0.
If they are connected incorrectly, the camera does not
work.
} For the connector of USB 3.0, the terminal may be blue or it is
printed as "SS" depending on controllers.
CAUTION
USB connector of USB 3.0
USB connector of USB 2.0