SDA Stand  

for Stereomicroscopes 

 

Operating Manual

 

Summary of Contents for Stemi 508

Page 1: ...SDA Stand for Stereomicroscopes Operating Manual...

Page 2: ......

Page 3: ...not covered by an update service Unless expressly authorized forwarding and duplication of this document as well as utilization and communication of its contents are not permitted Violations will ent...

Page 4: ...e SDA stand 9 3 2 Installing Stemi mount and microscope body 11 4 Operation 12 4 1 Vertical adjustment of the double boom 12 4 2 Longitudinal adjustment of the double boom 13 4 3 Boom rotation about t...

Page 5: ...warnings that must be observed by the user The following warning and information symbols are used in this manual CAUTION This symbol indicates a possible hazard to the user of the instrument CAUTION T...

Page 6: ...teps be taken to minimize damage If notified of such a defect the manufacturer is obligated to rectify it at his discretion either by repairing the instrument or delivering an intact replacement No gu...

Page 7: ...DESCRIPTION SDA Stand for Stereomicroscopes Carl Zeiss B 46 0075 e 05 06 7 2 DESCRIPTION 2 1 System overview...

Page 8: ...ment range of double boom Horizontal in guiding direction max 330 mm Vertical on SDA column 425 mm Rotary on SDA column 360 Accessible object positions Swivel radius around SDA column measured up to A...

Page 9: ...ols For the installation of the SDA stand remove all components from the packaging verify that all components are present CAUTION When unpacking and installing the base plate consider that it has a we...

Page 10: ...r position After releasing the lever it will automatically snap in again by a spring mechanism CAUTION The double boom has a weight of approx 15 kg call on a second person for help if necessary Slip t...

Page 11: ...With a short working distance of the supplementary lens system objective you may also slip the Stemi mount turned about over the column 32 to increase the spacing to the object and the base plate If...

Page 12: ...1 1 may only be released in turn otherwise the double boom will slip down unsecured involving the risk of injury to the operator or damage to the microscope and the object to be examined 4 1 Vertical...

Page 13: ...adjustment of the double boom CAUTION Double boom and weight of the microscope must be adjusted to each other to ensure stability of the setup and prevent the risk of injury to the operator or damage...

Page 14: ...and clamp it tightly 4 1 3 Loosen the boom rotation clamp screw 4 1 1 and swivel the double boom to the desired position CAUTION Take care to prevent any collision with persons the object to be examin...

Page 15: ...base plate Move the double boom up before if necessary Loosen clamp lever 4 1 5 on rotating tilting head 4 1 7 Incline column 32 with Stemi mount and microscope into the desired position Retighten the...

Page 16: ...sible For further details please contact Specifications are subject to change B 46 0075 e 05 06 Carl Zeiss MicroImaging GmbH P O B 4041 37030 G ttingen Germany Phone 49 0 551 5060 660 Fax 49 0 551 506...