DMTA017-01EN, Rev. F, August 2016
Maintenance
31
To replace a conventional ultrasonic probe
1.
Turn off the couplant flow.
2.
Disconnect the couplant tubes from the couplant source, or from the wedge.
3.
Disconnect the probe cable.
4.
Remove the scanner from the inspection surface.
5.
Using a flathead screwdriver, loosen the screw at the back of the spring-loaded
arm (SLA), just enough to free the yoke from the arm (see Figure 4-2 on page 31).
Figure 4
‑
2 The screw that secures the yoke to the SLA
6.
Remove the yoke.
7.
Using a hex socket-head screwdriver, remove the screw (on the side of the yoke)
that holds the parts together (see Figure 4-3 on page 32).