DMTA017-01EN, Rev. F, August 2016
Maintenance
35
Before installing the new probe into the probe holder, ensure that there is enough
couplant between the probe and the wedge.
10. Install and tighten the screws to secure the probe within the holder.
Do not overtighten the screws; doing so could crack the wedge.
11. Reinstall the yoke onto the holder.
12. Reinstall the yoke, probe holder, and probe assembly onto the spring-loaded arm.
This concludes the procedure for replacing a phased array probe.
4.4
Unit Cleaning
The HSMT-Compact scanner’s external surfaces can be cleaned when needed.
The magnetic wheels can be strongly attracted to any ferromagnetic materials or
surfaces, which potentially can cause injuries (for example, pinched fingers) or
equipment damage.
To clean the unit
1.
Ensure the unit is turned off by removing the source of power.
2.
Disconnect all cables.
3.
To bring the unit back to its original finish, clean the housing with a soft cloth.
4.
To remove persistent stains, use a damp cloth with a soft, soapy solution. Do not
use abrasive products or powerful solvents that could damage the finish.
NOTE
CAUTION