DMTA017-01EN, Rev. F, August 2016
Chapter 2
26
3.
Connect the water-manifold intake to the feed system (see Figure 2-10 on
page 25).
4.
Connect the water-manifold outlet to the wedge-couplant intake.
If only two out of the four water-manifold outlets are used, connect a tube
between the two unused outlets (see Figure 2-10 on page 25).
5.
Adjust the probe position as follows:
a)
Loosen the thumb screw that secures the probe-holder bracket to the scanner
frame (see Figure 2-11 on page 26).
Figure 2
‑
11 Adjusting the probe position
b)
Slide the probe holder to the desired position, and then tighten the thumb
screw.
For the TOFD ST1 and ST2 models exclusively, the edges of the probe-holder brackets
are in line with the beam exit point of the wedge to facilitate probe positioning.
Equipment connections and probe positions vary according to the application. To
mount the components, you might have to add or correctly position T-nuts in the
frame slots.
Thumb screw
Bracket edge
TIP
NOTE