Fig. 2
Operation Using the LED Lamp
Operation Using the Halogen Bulb
1. The surfaces of the lamp housing on the rear of the microscope will
become extremely hot during operation. When installing the microscope,
make sure to allow ample free space (10 cm or more) around and in
particular above the lamp housing for halogen bulb.
2. When installing the microscope, route the power cord away from the
lamp housing. Should the power cord come in contact with the hot lamp
housing, the power cord could melt and cause electric shock.
3. The standard service life of the lamp housing is eight (8) years of use or
20,000 hours of total power ON period, whichever is the shorter period.
For details, see Inspection Sheet on page 45.
4. To avoid potential shock hazards and burns when replacing the halogen
bulb, set the main switch
3
to “ ” (OFF) and allow the lamp housing
| and bulb to cool before touching them (Fig. 2).
Designated halogen bulb
12V100WHAL-L (PHILIPS 7724)
Safety note on LED (Light Emitting Diode)
The LED incorporated in the LED lamp housing is a class 1 LED
product. The light emitted by LED is basically safe, but do not look at
the illumination light directly for an extended period to prevent your eye
from being injured
CLASS 1 LED PRODUCT
4
3