Fig. 32
Adjusting the X-Axis/Y-Axis Knob
Rotation Tension
4
(Fig. 32)
} The rotation tension of the X-axis and Y-axis knobs can be adjusted
independently.
1. Loosen the 2 set screws
@
of a knob using the provided Allen wrench,
hold the stage so that it will not move, then turn the knob to adjust the
tension. Turning it in the direction of the arrow increases the tension and
turning in the opposite direction decreases the tension.
2. After adjustment, tighten the set screws firmly.
If the tension of a knob is too heavy or too light, skipping or
returning of image may occur during the stage movement.
CAUTION
CAUTION
CAUTION
Fig. 33
5
Using the Light Shielding Sheet
(Fig. 33)
The light shielding sheet provided with the reflected light flu-
orescence illuminator is too small to be used with the BX63.
Always use the light shielding sheet provided with the BX63.
} During fluorescence observation using a low-magnification objective,
the fluorescence image may be deteriorated due to light reflected from
the condenser or the surroundings. In this case, use the light shielding
sheet.
1. Lower the condenser to the lower limit position using the condenser
height adjustment knob.
2. Insert the light shied sheet
@
all the way into the gap between the upper
and lower stages on the side of the stage (IX-SVL2).
If the condenser is lowered insufficiently, the sheet cannot be
inserted into the normal position and the light shielding effect
cannot be obtained.