21
BX51WI
Fig. 32
Fig. 33
Fig. 34
With the WI-UCD Condenser (Figs. 32-34)
}When selecting the brightfield (BF) light path using the WI-UCD, leave
one DIC prism (large) mount position empty.
1. Remove the WI-UCD condenser from the microscope frame.
2. Remove the condenser cover @ by loosening the retaining screws ²
using a coin, etc.
3. Attach the suitable DIC prism for the objective in use as described below.
· Using the knob provided with the condenser, loosen the two DIC prism
clamping screws ³ until the rotatable limits.
· Rotate the turret by 90° counterclockwise, and drop in the DIC prism by
aligning its positioning pin ƒ with the positioning groove … in the hole of
the turret |. (Fig. 33)
# Be careful not to touch the prism inside the frame.
A. Rotate the turret | by 90° clockwise (Fig. 34) and tighten the two DIC
prism clamping screws ³ uniformly using the dedicated knob provided
with the condenser. (Figs. 33 & 34)
# Do not tighten the screws too much, or the prism frame may be
deformed.
B. Rotate the turret | by 90° clockwise (Fig. 34), and attach the index sticker
† provided with the DIC prism onto the side ‡ of the condenser turret |
so that the index sticker is upside down. (Figs. 33 & 34)
4. After attaching all of the required DIC prisms, attach the cover @ and
tighten the retaining screws ². (Fig. 32)
5. Attach the condenser back onto the microscope frame.
@
²
³
|
†
‡
…
ƒ
A
B