94MS36 Rev. F2 S/W Ver 2.06
2-5
•
Press the
P
key on the integrator to power up the lights, the front
switches, and the printer's computer system.
Note:
Allow the printer a 5 minute warm-up period before any
calibration procedures. This will allow for complete and accurate
light level stabilization.
•
Press the integrator
I
key to select the desired light level to be calibrated,
always start at Low Intensity.
For 8kW
Low 2000 W
Medium 5000 W
High 8000 W
For 6kW
Low 2000 W
Medium 3500 W
High 6000 W
For 5kW
Low 1000 W
Medium 2500 W
High 5000 W
•
In order to prevent possible damage to the high resolution printers Mylar®
while calibrating we recommend you set the no vacuum feature.
•
Press the
V
key on the keypad, the Memory and Exposure windows
will dim.
•
Press the
9
key three times.
•
Press the
V
key once. This will prevent vacuum while calibrating. The
Vacuum window will read
[_---]
.
•
The high resolution printer's calibration has been set at the factory.
[Ut]
Upper frame top surface
[Ub]
Upper frame bottom surface
[Lt]
Lower frame top surface
[Lb]
Lower frame bottom surface
Note:
Preset all 12 calibration factor numbers to 500 units. Place a
slim UV meter facing up on the center of the top frame, below
the Mylar®.
•
Enter a 500 unit exposure into the integrator. Be sure the seconds mode
is off.
•
Press the
G
button twice. This will send the frame into the unit and
start the exposure.
•
After the exposure, retrieve your frame by holding down the
S
button
and press
G
button. When the frames start to move, release both
buttons.
•
Record the resulting mj reading from the meter.