9
– Maintenance
79
Chapter 9
| Maintenance
Maintenance mainly involves replacing any sensors that no longer meet their
original metrological specifications.
The operations explained in this section must be performed by
authorized, qualified personnel because they could affect detection
reliability. Inspection and maintenance must be carried out in
accordance with the current editions of EN60079-17 and eventually
with any local or national additional requirements that may apply in
the country of installation.
Possible transmitter errors
The table below lists various potential detector errors.
Fault observed
Possible cause
Action
Page
0 mA line current
Connection cable
Check the cable.
-
Power supply
Check the voltage at the
transmitter's terminals (see
Alim V
under the
Maintenance
menu).
Electronic board
Replace the board.
-
Line current
> 0 mA and < 1mA
Sensor
Replace the sensor (see
Sensor Replacement
menu).
Line resistance too high
Check the cable.
Power supply
Check the voltage at the
transmitter's terminals (see
Alim V
under the
Maintenance
menu).
Improper calibration gas
Check the concentration of
the calibration gas
-
Check the input value (see
Calibration gas
under
Sensor
Replacement
menu)
Zeroing not
possible
Sensor
Replace the sensor (see
Sensor Replacement
menu).
Sensitivity
adjustment
impossible
Sensor
Replace the sensor (see
Sensor Replacement
menu).
"
SUP
" displayed
Verification required
Acknowledge verification.
Summary of Contents for olct 80
Page 1: ...1 User manual OLCT 80 Part Number NPO80GB Revision E 1 The Fixed Gas Detection Experts...
Page 6: ...Table of Contents vi...
Page 10: ...1 Overwiew 4...
Page 18: ...2 Transmitter Overview 12...
Page 22: ...3 Menus 16 Figure 15 the OLCT 80 s sub menus under AFF MES DATE TIME and PROG...
Page 23: ...3 Menus 17 Figure 16 the OLCT 80 s sub menus under MAINT CALIBRA 4 20mA INFOS and TEST...
Page 42: ...3 Menus 36 Figure 26 RELAY 1 menu Press ESC repeatedly to return to the reading display...
Page 58: ...3 Menus 52...
Page 74: ...5 Wireless Version 68...
Page 82: ...7 Pre installation Hardware Configuration 76...
Page 84: ...8 Preventative Maintenance 78...
Page 90: ...10 Accessories 84...
Page 96: ...12 EU Declarations of Conformity 90 OLCT 80 without antenna...
Page 97: ...12 EU Declarations of Conformity 91...
Page 98: ...12 EU Declarations of Conformity 92 OLCT 80 with antenna...
Page 99: ...12 EU Declarations of Conformity 93...
Page 100: ...12 EU Declarations of Conformity 94 IR20 remote control of the OLCT 80...
Page 108: ...13 Technical Specifications 102...
Page 116: ...14 Special instructions for use in explosive environments and fonctional safety 110...
Page 119: ...113...