8
– Preventative Maintenance
77
Chapter 8
| Preventative Maintenance
Periodic inspections ensure that the equipment and system is functioning
properly and providing reliable detection services. The section describes the
preventative maintenance procedures required and how often they are to be
performed. Inspection and maintenance must be carried out in accordance with
the current editions of EN60079-17 and eventually with any local or national
additional requirements that may apply in the country of installation.
Maintenance frequency
Gas detectors are safety devices. Oldham recommends regular testing of fixed
gas detection installations. This type of test involves injecting a standard gas of
sufficient concentration into the detector to trigger pre-set alarms. This test does
not, in any event, replace a full calibration of the detector.
Frequency of gas testing depends on the industrial application in which the
detector is used. Inspection should occur frequently during the months following
installation start-up; later it may be spaced out if no significant problem is
observed. If a detector does not react upon contact with gas, it must be
calibrated. Calibration frequency will depend on the results of these tests
(moisture, temperature, dust, etc.); however, the device should be calibrated at
least once per year.
The site manager is responsible for implementing the safety procedures at the
site. Oldham is not responsible for implementing safety procedures.
OLCT 80
Periodic maintenance involves the following steps:
■
Remove dust from the sensor's protective housing, using a dry cloth only.
Do not use water or any type of solvent.
■
When using the equipment in dusty explosive atmospheres, the equipment
should be thoroughly cleaned on a regular basis to prevent the build-up of
dust. If a layer of dust does build up on the detector, this layer may not
exceed 5 mm.
■
Replace the screws: use high-quality screws > A4.70.
■
Perform the zero test with clean air: follow the steps described under
Sensor
calibration
on page 45 in the event of deviation.
■
Perform the gas sensitivity test: follow the steps described under
Sensor
calibration
on page 45 in the event of deviation.
Summary of Contents for olct 80
Page 1: ...1 User manual OLCT 80 Part Number NPO80GB Revision E 1 The Fixed Gas Detection Experts...
Page 6: ...Table of Contents vi...
Page 10: ...1 Overwiew 4...
Page 18: ...2 Transmitter Overview 12...
Page 22: ...3 Menus 16 Figure 15 the OLCT 80 s sub menus under AFF MES DATE TIME and PROG...
Page 23: ...3 Menus 17 Figure 16 the OLCT 80 s sub menus under MAINT CALIBRA 4 20mA INFOS and TEST...
Page 42: ...3 Menus 36 Figure 26 RELAY 1 menu Press ESC repeatedly to return to the reading display...
Page 58: ...3 Menus 52...
Page 74: ...5 Wireless Version 68...
Page 82: ...7 Pre installation Hardware Configuration 76...
Page 84: ...8 Preventative Maintenance 78...
Page 90: ...10 Accessories 84...
Page 96: ...12 EU Declarations of Conformity 90 OLCT 80 without antenna...
Page 97: ...12 EU Declarations of Conformity 91...
Page 98: ...12 EU Declarations of Conformity 92 OLCT 80 with antenna...
Page 99: ...12 EU Declarations of Conformity 93...
Page 100: ...12 EU Declarations of Conformity 94 IR20 remote control of the OLCT 80...
Page 108: ...13 Technical Specifications 102...
Page 116: ...14 Special instructions for use in explosive environments and fonctional safety 110...
Page 119: ...113...