background image

 

17

 

3

3

.

.

 

 

T

T

E

E

S

S

T

T

I

I

N

N

G

G

 

 

o

o

f

f

 

 

r

r

e

e

l

l

a

a

y

y

s

s

 

 

a

a

n

n

d

d

 

 

e

e

x

x

t

t

e

e

r

r

n

n

a

a

l

l

 

 

c

c

o

o

n

n

t

t

r

r

o

o

l

l

s

s

 

 

It  is  also  recommended  to  periodically  check  alarm  activation  and  the  corresponding  external 
controls. 
 
 

 

4

4

.

.

 

 

C

C

l

l

e

e

a

a

n

n

i

i

n

n

g

g

 

 

Do not use alcohol or ammonia-based liquids to clean the central unit. 
If needed, clean the outside of the case with a damp cloth. 

 
 
 

5

5

.

.

 

 

C

C

e

e

n

n

t

t

r

r

a

a

l

l

 

 

u

u

n

n

i

i

t

t

 

 

m

m

a

a

i

i

n

n

t

t

e

e

n

n

a

a

n

n

c

c

e

e

 

 

 

 

The central unit does not require special maintenance. 
If sensor calibration is required, this operation should only be performed by qualified personnel. 

 

Gas  detection  instruments  are  potential  life-saving  devices.  Recognizing  this  fact,  Industrial 
Scientific Corporation recommends that a functional “bump” test be performed on every fixed gas-
monitoring instruments as part of a regular maintenance program. A functional test is defined as a 
brief exposure of the detector to a concentration of gas(es) in excess of the lowest alarm set-point 
for each sensor for the purpose of verifying sensor and alarm operation and is not intended to be a 
measure of the accuracy of the instrument. 
Industrial  scientific  further  recommends  that  a  full  instrument  calibration  be  performed  using  a 
certified  concentration(s)  of  calibration  gas(es)  quarterly,  every  3  months.*  Calibrations  may  be 
necessary more or less frequently based, for example, on application, field conditions, exposure to 
gas,  sensor  technology,  and  environmental  conditions.  The  frequency  of  calibration  is  best 
determined by company policy or local regulatory agencies. 
If  an  instrument  fails  to  operate  properly  during  any  functional  “bump”  test,  a  full  instrument 
calibration should be performed successfully prior to use. 
These recommendations are based on safe work procedures, industry best practises, and regulatory 
standards to ensure worker safety. Industrial scientific is not responsible for setting safety practices 
and policies. 

 

*  For  new  installations  it  may  be  prudent  to  carry  out  bump  tests  frequently  at  first  (perhaps 
weekly),  increasing  the  time  intervals  (to,  perhaps,  monthly  or  more)  as  confidence  grows  with 
experience in the installation concerned, on the basis of the maintenance record. 

 

 

REFER TO CHAPTERS 4.2.5 AND 4.2.8 (INITIALISATION AND TEST) 

 

Do  not  forget  to  initialise  the  system  upon  cell  replacement:  use  the  “INI”  menu 
(chapter 4.2.5). 

 

Summary of Contents for MX15

Page 1: ...I IN NS ST TA AL LL LA AT TI IO ON N A AN ND D O OP PE ER RA AT TI IN NG G I IN NS ST TR RU UC CT TI IO ON NS S Ref NPM15GB...

Page 2: ...uired particularly for electrical installation and assembly connections service lines Any modification of the equipment as well as use of unauthorized parts will result in cancellation of any warranty...

Page 3: ...ECIFIC USE C C L L A A U U S S E E S S C C O O N N C C E E R R N N I I N N G G P P R R O O P P E E R R T T Y Y Drawings plans specifications and information included in this document contain confident...

Page 4: ...ly 230 Volts 115VAC on request 11 2 2 2 Power supply 24 VDC protection through F8 can be connected on terminals 0 and 24V as indicated in the wiring examples at the end of the manual 11 2 3 Measuring...

Page 5: ...otentiometers A motherboard including all system components power supply display relays and connectors A front face with tactile keys I I I I T T e e c c h h n n i i c c a a l l s s p p e e c c i i f...

Page 6: ...ammable Display front Type LCD screen 4 digits 7 segments 3 characters 14 icon segments 4 LEDs Description of unit and gas programmable by user in a list 3 characters edited by user Keyboard through t...

Page 7: ...h information provided to external units The site manager where the equipment is installed should take into consideration and comply with the information in the following paragraphs Refer to the provi...

Page 8: ...l unit in configuration 0 at 100 LEL The following curve provides a response to the central unit regarding the measured value and fault processing based on the value of the input current supplied by t...

Page 9: ...n attended area control centre console room control room etc and in a less humid no condensation and temperate environment see chapters 2 1 and 3 2 2 C Ce en nt tr ra al l u un ni it t e el le ec ct t...

Page 10: ...should be wired on the two terminals L orange and N blue as indicated in the wiring examples at the end of manual 2 2 2 Power supply 24 VDC protection through F8 can be connected on terminals 0 and 2...

Page 11: ...e in positive safety negative upon request and potential free REL1 terminals match the relay 1 contacts alarm 1 REL2 terminals match the relay 2 contacts alarm 2 The relay contacts can be used normall...

Page 12: ...replaced with an interrupted line on the display unit To return to the measurement display press one of the two keys or To return to normal operation simultaneously press the and keys 2 2 M Me en nu u...

Page 13: ...larms 1 manual MAN if the gas concentration again becomes lower than the preset alarm threshold this alarm must be acknowledged manually by pressing the acknowledgment key 2 or automatical AUT if the...

Page 14: ...changes the access code for various menus 1000 in standard upon shipment of equipment display the CODE menu and key validate it enter display current code indicate the new code using the and keys vali...

Page 15: ...s concentration is correct by reading the display unit Otherwise adjust the sensitivity using a S potentiometer Disconnect the standard gas cylinder Wait for the return to zero on the display unit Pre...

Page 16: ...intended to be a measure of the accuracy of the instrument Industrial scientific further recommends that a full instrument calibration be performed using a certified concentration s of calibration ga...

Page 17: ...onal use of natural resources MX15 has to be the object of a selective collection for the electronic equipments and cannot be scrapped with the normal domestic waste The user thus has the obligation t...

Page 18: ...gment 1 3 relay contact outputs potential free NO or NF based on position of jumpers next to relays DEF RL1 RL2 Connector of first sensor with 2 wires Example of connection between a MX15 central unit...

Page 19: ...0 TWIN type sensors L N T MAINS 230VAC 0 24V Remote acknowledgment 1 2 3 Connector of first sensor with 3 active wires relay contact outputs potential free NO or NF based on position of jumpers next t...

Page 20: ...ring 1 2 3 1 2 3 Example of connection with an EXPLO or TOXIC sensor with 3 active wires L N T MAINS 230VAC 0 24V relay contact outputs potential free NO or NF based on position of jumpers next to rel...

Page 21: ...22...

Page 22: ...23 Ex Ox Tox Gasdetectie Westerdreef 5V 2152 CS Nieuw Vennep Telefoon 0252 620885 E mail info exoxtox n Website www exoxtox nl...

Reviews: